FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3182299 · Received June 21, 2013

Report

Report Number
1416980-2013-16147
Event Type
Malfunction
Date Received
June 21, 2013
Date of Event
May 30, 2012
Report Date
May 31, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

A HIGH DRAIN ERROR 104 (NIGHT DRAIN #4) ALARM WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. A HIGH DRAIN ALARM IS INDICATIVE OF AN IIPV SITUATION. THIS ALARM INDICATES THE PATIENT DRAINED GREATER THAN 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME IN STANDARD MODE. THE ALARM OCCURRED ON (B)(6) 2012 AT 21:47:38. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
282982 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 41 YR