FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3182290 · Received June 21, 2013

Report

Report Number
2531779-2013-08735
Event Type
Malfunction
Date Received
June 21, 2013
Report Date
May 26, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 08/14/2013 - DEVICE EVALUATION:THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/22/2013 WITH THE FOLLOWING FINDINGS:DURING TESTING, THE PUMP WAS POWERED ON AND THE DISPLAY WAS BLANK. DURING INVESTIGATION THE PUMP CASE WAS OPENED AND THE DISPLAY SCREEN WAS FOUND TO BE CRACKED. A TEST SCREEN WAS INSERTED AND WAS FOUND TO FUNCTION PROPERLY.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THE REPORTER STATED THAT THE DISPLAY SCREEN HAD BECOME DIM AND DISCOLORED AFTER THE PATIENT FELL WHILE WEARING THE PUMP. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
282803 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 11 YR