FDA Adverse Event Injury Summary report: N

XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 3182284 · Received June 21, 2013

Report

Report Number
2024168-2013-03856
Event Type
Injury
Date Received
June 21, 2013
Date of Event
May 28, 2013
Report Date
May 29, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE STENT DISLODGEMENT WAS CONFIRMED. THE FAILURE TO ADVANCE/CROSS COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. THE DIFFICULTY TO REMOVE (RESISTANCE WITH THE GUIDING CATHETER) COULD NOT BE TESTED DUE TO THE CONDITION OF THE RETURNED DEVICE. BASED ON VISUAL AND DIMENSIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT USING A FEMORAL ARTERY ACCESS APPROACH DURING A PROCEDURE OF THE MODERATELY TORTUOUS, MODERATELY CALCIFIED LEFT ANTERIOR DESCENDING (LAD) ARTERY AFTER PREDILATATION THE 2.75 X 28 MM XIENCE XPEDITION STENT DELIVERY SYSTEM (SDS) WAS ADVANCED BUT FAILED TO CROSS THE LESION. DURING RETRACTION AND INTRODUCING INTO THE GUIDE CATHETER THE STENT IMPLANT BECAME DISLODGED OFF THE BALLOON INTO THE ANATOMY. THE STENT IMPLANT WAS REMOVED BY USING A SNARE DEVICE. THERE WAS NO REPORTED ADVERSE PATIENT EFFECT. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. AFTER ADDITIONAL PREDILATATION, A SECOND 2.75 X 28 MM XIENCE XPEDITION STENT WAS IMPLANTED WITHOUT ISSUE. THERE WAS NO REPORTED ADVERSE PATIENT EFFECT. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
282801 XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 2082261

Patients

Seq Age Sex Outcome Treatment
1 73 YR Required Intervention