ENDOTRACHEAL TUBE 8229307 NIM EMG 7MM RE
Report
- Report Number
- 1045254-2013-00429
- Event Type
- Injury
- Date Received
- June 21, 2013
- Date of Event
- May 23, 2013
- Report Date
- May 23, 2013
- Manufacturer
- MEDTRONIC XOMED, INC.
- Product Code
- ETN
- PMA / PMN Number
- K925640
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
Narratives
THIS DEVICE IS USED FOR THERAPEUTIC PURPOSES. (B)(4): THE DEVICE WAS RETURNED FOR EVALUATION. THE PRODUCT ANALYSIS FOUND THE TUBING AND ELECTRODE WIRES ARE BENT; THE BLUE ELECTRODE WIRES ARE OUT OF THE LUMEN UNDER THE CUFF AND 1 HAD PUNCTURED THE CUFF. THE INSTRUCTIONS FOR USE [68E4127 REVISION B] INDICATE: ¿DO NOT ATTEMPT TO USE AN EMG TUBE WITHOUT FIRST PERFORMING A CUFF INFLATION TEST¿ THIS WOULD HAVE DETECTED A PUNCTURED OF THE CUFF PRIOR TO USE. THE PRODUCT EVALUATION FINDINGS ARE INDICATIVE OF DAMAGE TO THE TUBE/CUFF/ELECTRODE WIRES AS A RESULT OF DEVICE MISHANDLING.
IT WAS REPORTED THAT DURING A PROCEDURE ¿THERE WAS A PROBLEM WITH THE TUBE AND THE CUFF KEPT DEFLATING. THROAT PACKS WERE USED AND THE TUBE WAS REPLACED AT A SAFE AND APPROPRIATE TIME DURING THE PROCEDURE.¿ THE SURGERY WAS COMPLETED WITHOUT FURTHER INCIDENT. THERE WAS NO REPORT OF PATIENT IMPACT OR INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 282229 | ENDOTRACHEAL TUBE 8229307 NIM EMG 7MM RE | STIMULATOR, NERVE | ETN | MEDTRONIC XOMED, INC. | 8229307 | 0206398997 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00051 YR | Required Intervention |