COBAS 6000 C501 MODULE
Report
- Report Number
- 1823260-2013-03758
- Event Type
- Malfunction
- Date Received
- June 21, 2013
- Date of Event
- June 4, 2013
- Report Date
- July 3, 2013
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JGS
- PMA / PMN Number
- K060373
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA.
BASED ON THE CALIBRATION INFORMATION PROVIDED, IT CAN BE SEEN THAT ONE OF THE STANDARDS USED FOR THE CALIBRATION WAS HIGHER CONCENTRATED. THE CUSTOMER MAY HAVE USED A HIGH STANDARD THAT HAD BEEN OPENED FOR TOO LONG. THE EFFECT IS SEEN IN THE COMPLAINED LOW SAMPLE RESULTS. THE ISSUE WAS RESOLVED ON SUBSEQUENT CALIBRATION.
THE CUSTOMER REPORTED THAT THEY RECEIVED QUESTIONABLE RESULTS FOR A TOTAL OF 50 PATIENT SAMPLES TESTED FOR ION SELECTIVE ELECTRODE (ISE) SODIUM. THE SAMPLES WERE INITIALLY RUN BETWEEN (B)(6) 2013. THE CUSTOMER NOTED THAT THE CALIBRATION FOR SODIUM AND CHLORIDE HAD COMPENSATED VALUES WHICH WERE NOT ACCEPTABLE ACCORDING TO THEIR LABORATORY PROCEDURES. QUALITY CONTROLS SHIFTED LOW, BUT WERE NOT OUTSIDE OF RANGE. PATIENT VALUES ALSO WENT LOWER. OF THE FIFTY QUESTIONED PATIENT SAMPLES, THE CUSTOMER PROVIDED EXAMPLES FOR FIVE PATIENT SAMPLES WHICH HAD ERRONEOUS RESULTS. ALL INITIAL SAMPLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY WHERE THEY WERE QUESTIONED BY THE NURSE MANAGER. SAMPLES WERE REPEATED ON AN INTEGRA 400 ANALYZER (SERIAL (B)(4). REPEAT VALUES WERE BELIEVED TO BE CORRECT AND CORRECTED REPORTS WERE ISSUED FOR THESE SAMPLES. SAMPLE ONE INITIALLY RESULTED AS 131 MMOL/L AND REPEATED AS 139 MMOL/L. SAMPLE TWO, FROM A MALE BORN ON (B)(6) 1947, INITIALLY RESULTED AS 127 MMOL/L AND REPEATED AS 136 MMOL/L. SAMPLE THREE, FROM A FEMALE BORN ON (B)(6)1981, INITIALLY RESULTED AS 130 MMOL/L AND REPEATED AS 138 MMOL/L. SAMPLE FOUR, FROM A FEMALE BORN ON (B)(6)1953, INITIALLY RESULTED AS 134 MMOL/L AND REPEATED AS 141 MMOL/L. SAMPLE FIVE, FROM A MALE BORN ON (B)(6) 1947, INITIALLY RESULTED AS 123 MMOL/L AND REPEATED AS 130 MMOL/L. THE PATIENTS WERE NOT ADVERSELY AFFECTED BY THE EVENT. THE LOT NUMBER AND EXPIRATION DATE OF THE SODIUM ELECTRODE WERE NOT PROVIDED. ON (B)(6) 2013, THE CUSTOMER REPLACED ALL ELECTRODES, ISE REAGENTS, PINCH VALVE, AND SIPPER TUBING. AFTER DOING THIS, THE CUSTOMER PRIMED THE ANALYZER, RAN THE GREEN CLEANING RACK, AND PERFORMED AN ISE CHECK. THE CUSTOMER THEN CALIBRATED AND RAN QUALITY CONTROLS SUCCESSFULLY. ON THE EVENING OF (B)(6) 2013, A TECHNOLOGIST CALIBRATED AND RAN QUALITY CONTROLS FOR THE ISE TESTS AND THESE WERE NOT ACCEPTABLE. CALIBRATION AND CONTROLS WERE RUN AGAIN AND THESE ALSO WERE NOT ACCEPTABLE. THE CUSTOMER THEN RAN THE GREEN CLEANING RACK AND PERFORMED AND ISE CHECK. AFTER DOING THIS, THE CUSTOMER REPEATED CALIBRATION AND QUALITY CONTROLS SUCCESSFULLY. CALIBRATION AND CONTROLS WERE RUN AGAIN ON THE MORNING OF (B)(6) 2013 AND THESE WERE ACCEPTABLE. THE FIELD SERVICE REPRESENTATIVE COULD NOT DETERMINE A CAUSE. HE CHECKED THE ISE FLUIDICS AND STATED THAT THE CUSTOMER REPLACED ALL ELECTRODES AND REAGENTS PRIOR TO HIS ARRIVAL. HE PERFORMED A PRECISION STUDY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 283836 | COBAS 6000 C501 MODULE | CLINICAL CHEMISTRY ANALYZER | JGS | ROCHE DIAGNOSTICS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 081 YR |