FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 3182188 · Received June 14, 2013

Report

Report Number
1627487-2013-04769
Event Type
Injury
Date Received
June 14, 2013
Date of Event
December 22, 2012
Report Date
May 22, 2013
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. CORRECTIVE AND PREVENTIVE ACTION (CAPA). INVESTIGATION WAS PERFORMED. POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 3 OF 3. REF MFR REPORT: 1627487-2013-04767 AND 1627487-2013-04768. IT WAS REPORTED THE PT HAD RECEIVED A BURN WHILE RECHARGING THE IPG WITH THE CHARGING SYSTEM. THE PT REPORTED THE INCIDENT HAD HAPPENED 5 MONTHS BEFORE AND A SCAB FORMED AND MORE RECENTLY THERE WAS A SCAR. THE PT REPORTED THE IPG WAS SHALLOW UNDER THE SKIN. IT WAS ALSO REPORTED THE SYSTEM WOULD TURN ON RANDOMLY AND A SHOCKING SENSATION WOULD OCCUR IN HER LEFT LEG WHICH WOULD CAUSE HER TO DRAW HER LEGS TOWARD HER CHEST. THE PT WAS NO LONGER RECEIVING EFFECTIVE STIMULATION AND HAD DISCONTINUED THE USE OF THE SCS SYSTEM. IT WAS REPORTED THE PHYSICIAN PLANNED TO EXPLANT THE SYSTEM. ON 08/01/2012, ST JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
270758 CHARGING SYSTEM SCS CHARGING SYSTEM LGW ST. JUDE MEDICAL - NEUROMODULATION 3721 3213352

Patients

Seq Age Sex Outcome Treatment
1 50 YR Other