EON MINI
Report
- Report Number
- 1627487-2013-13907
- Event Type
- Injury
- Date Received
- June 14, 2013
- Date of Event
- May 20, 2013
- Report Date
- May 20, 2013
- Manufacturer
- ST. JUDE MEDICAL, NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS IPG MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REFERENCE MFR. REPORT: 1627487-2013-13908. IT WAS REPORTED THE PATIENT EXPERIENCED POCKET HEATING WHILE CHARGING HIS IPG. THE PATIENT STATED IT LEFT A RED MARK ON HIS SKIN. THE PATIENT WAS GIVEN A NEW LE CHARGER AND A CHARGER BOX. ON 08/01/2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 270962 | EON MINI | SCS IPG | LGW | ST. JUDE MEDICAL, NEUROMODULATION | 3788 | 3632984 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Other | SCS LEAD: MODEL 3163| IMPLANT DATE: |