FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES

MDR report key: 3182103 · Received June 21, 2013

Report

Report Number
3015876-2013-00535
Event Type
Malfunction
Date Received
June 21, 2013
Date of Event
May 27, 2013
Report Date
May 27, 2013
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K102972
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE POWER PCB ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE.

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL FURTHER EVALUATED THE REMOVED POWER PCB ASSEMBLY AND DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO A SHORTED CAPACITOR, DESIGNATOR C4. THE SHORTED CAPACITOR CAUSED AN OPEN LAND ON THE PCB ASSEMBLY BETWEEN CAPACITORS C25 AND C4.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEIR DEVICE WOULD NO LONGER POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
283503 LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES DEFIBRILLATORS, AUTOMATIC, EXTERNAL MKJ PHYSIO-CONTROL, INC

Patients

Seq Age Sex Outcome Treatment
1