FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES
MDR report key: 3182103
·
Received June 21, 2013
Report
- Report Number
- 3015876-2013-00535
- Event Type
- Malfunction
- Date Received
- June 21, 2013
- Date of Event
- May 27, 2013
- Report Date
- May 27, 2013
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K102972
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4): PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE POWER PCB ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE.
Additional Manufacturer Narrative · 1
PHYSIO-CONTROL FURTHER EVALUATED THE REMOVED POWER PCB ASSEMBLY AND DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO A SHORTED CAPACITOR, DESIGNATOR C4. THE SHORTED CAPACITOR CAUSED AN OPEN LAND ON THE PCB ASSEMBLY BETWEEN CAPACITORS C25 AND C4.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THEIR DEVICE WOULD NO LONGER POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.