FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3181933 · Received June 17, 2013

Report

Report Number
8020893-2013-01328
Event Type
Malfunction
Date Received
June 17, 2013
Date of Event
May 23, 2013
Report Date
May 23, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN...
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT AN 840 VENTILATOR STOPPED CYCLING. NO PT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION, THE CSE INSPECTED THE DEVICE AND REPLACED THE BREATH DELIVERY UNIT (BDU) CPU PCB. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
273453 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR PURITAN... 840

Patients

Seq Age Sex Outcome Treatment
1