FDA Adverse Event
Malfunction
Summary report: N
LL100 CRYOSURGICAL SYSTEM
MDR report key: 3181926
·
Received June 17, 2013
Report
- Report Number
- 1216677-2013-00018
- Event Type
- Malfunction
- Date Received
- June 17, 2013
- Date of Event
- April 24, 2013
- Report Date
- June 17, 2013
- Manufacturer
- COOPERSURGICAL, INC.
- Product Code
- GEH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE RETURNED LL100 CRYO GUN WAS THOROUGHLY INVESTIGATED BY OUR SVC AND REPAIR DEPT AND THE REPORTED CONDITION COULD NOT BE CONFIRMED. THE UNIT TESTED TO SPECIFICATION AND THE TIP WOULD NOT COME OFF. PLEASE NOTE - THE ONLY WAY THE TIP COULD COME OFF IS IF IT WAS NOT SCREWED ONTO THE SHAFT PROPERLY. THE THREADS WERE NOTED TO BE IN PERFECT CONDITION. (B)(4).
Description of Event or Problem · 1
DOCTOR WAS DEMONSTRATING THE CRYO GUN AND THE TIP BLEW OFF - IT DID NOT IMPACT ANYONE. THERE WAS NO PT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 274336 | LL100 CRYOSURGICAL SYSTEM | CRYO GUN | GEH | COOPERSURGICAL, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |