FDA Adverse Event Malfunction Summary report: N

LL100 CRYOSURGICAL SYSTEM

MDR report key: 3181926 · Received June 17, 2013

Report

Report Number
1216677-2013-00018
Event Type
Malfunction
Date Received
June 17, 2013
Date of Event
April 24, 2013
Report Date
June 17, 2013
Manufacturer
COOPERSURGICAL, INC.
Product Code
GEH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE RETURNED LL100 CRYO GUN WAS THOROUGHLY INVESTIGATED BY OUR SVC AND REPAIR DEPT AND THE REPORTED CONDITION COULD NOT BE CONFIRMED. THE UNIT TESTED TO SPECIFICATION AND THE TIP WOULD NOT COME OFF. PLEASE NOTE - THE ONLY WAY THE TIP COULD COME OFF IS IF IT WAS NOT SCREWED ONTO THE SHAFT PROPERLY. THE THREADS WERE NOTED TO BE IN PERFECT CONDITION. (B)(4).

Description of Event or Problem · 1

DOCTOR WAS DEMONSTRATING THE CRYO GUN AND THE TIP BLEW OFF - IT DID NOT IMPACT ANYONE. THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
274336 LL100 CRYOSURGICAL SYSTEM CRYO GUN GEH COOPERSURGICAL, INC.

Patients

Seq Age Sex Outcome Treatment