FDA Adverse Event Malfunction Summary report: N

ORTHO SUMMIT SAMPLE HANDLING SYSTEM

MDR report key: 3181909 · Received June 21, 2013

Report

Report Number
2250051-2013-00170
Event Type
Malfunction
Date Received
June 21, 2013
Date of Event
May 28, 2013
Report Date
June 21, 2013
Manufacturer
HAMILTON BONADUZ AG
Product Code
MMH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2013 AN OCD FIELD ENGINEER (FE) ARRIVED AT THE CUSTOMER SITE AND FOUND THE PLUNGER IN POSITION # 3 AND 10 WERE FAULTY. THE PLUNGER CLAMPS WERE REPLACED. THE FE TESTED THE SUMMIT LLD WITH SPLLD AND (B)(4) SOFTWARE. ALL TEST PASSED WITHOUT ERROR. REPAIRS HAVE RETURNED THIS INSTRUMENT TO EXPECTED OPERATION

Description of Event or Problem · 1

THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT DID NOT PIPETTE THE CORRECT AMOUNT OF SAMPLE AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
282084 ORTHO SUMMIT SAMPLE HANDLING SYSTEM SAMPLE PIPETTOR MMH HAMILTON BONADUZ AG 934600

Patients

Seq Age Sex Outcome Treatment