HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2013-16125
- Event Type
- Malfunction
- Date Received
- June 21, 2013
- Date of Event
- May 31, 2013
- Report Date
- May 31, 2013
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
(B)(4).AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. THIS WAS A REPORT OF A SYSTEM ERROR 2240 WHERE HOME PATIENT LEFT THE CLAMPS OPEN ON UNUSED LINES. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE WARNS THE USER TO MAKE SURE ALL CLAMPS ON UNUSED FLUID LINES ARE CLOSED SECURELY AND INSTRUCTS THE USER TO CLOSE ALL CLAMPS WHILE PREPARING TO LOAD THE DISPOSABLE SET. THE GUIDE ALSO WARNS THE USER THAT A SYSTEM ERROR 2240 CAN BE CAUSED BY UNCLAMPED, UNUSED SUPPLY LINES. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT WHILE PERFORMING PERITONEAL DIALYSIS, A HOME PATIENT (HP) HAD RECEIVED A SYSTEM ERROR 2240 (AIR IN SET) ALARM ON A HOMECHOICE (HC) MACHINE DURING DWELL THREE OF FOUR. THE HP HAD LEFT CLAMPS OPEN ON UNUSED LINES. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HP IN CLEARING THE ALARM. THE HP WAS TO FINISH THERAPY WITH MANUAL SUPPLIES. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED IN ASSOCIATION WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 282916 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 84 YR | HOMECHOICE |