FDA Adverse Event Malfunction Summary report: N

XIENCE PRIME EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 3181886 · Received June 21, 2013

Report

Report Number
2024168-2013-03849
Event Type
Malfunction
Date Received
June 21, 2013
Date of Event
May 20, 2013
Report Date
May 27, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4) - EXCESSIVE FORCE, NO PRE-DILATATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE SHAFT SEPARATION WAS CONFIRMED. THE FAILURE TO ADVANCE/CROSS COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. BASED ON VISUAL AND DIMENSIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. IT SHOULD BE NOTED THAT THE XIENCE PRIME EVEROLIMUS ELUTING CORONARY STENT SYSTEM INSTRUCTIONS FOR USE (IFU) STATES: PRE-DILATE THE LESION WITH A PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA) CATHETER. ADDITIONALLY, THE IFU STATES: SHOULD ANY RESISTANCE BE FELT AT ANY TIME DURING EITHER LESION ACCESS OR REMOVAL OF THE DELIVERY SYSTEM POST-STENT IMPLANTATION, THE ENTIRE SYSTEM SHOULD BE REMOVED AS A SINGLE UNIT AND THAT APPLYING EXCESSIVE FORCE TO THE DELIVERY SYSTEM CAN POTENTIALLY RESULT IN LOSS OR DAMAGE TO THE STENT AND / OR DELIVERY SYSTEM COMPONENTS. IN THIS CASE, THE IFU DEVIATION(S) APPEAR TO HAVE CONTRIBUTED TO THE REPORTED DIFFICULTIES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION LOCATED IN THE MILDLY TORTUOUS MID RIGHT CORONARY ARTERY. A GUIDE WIRE WAS CROSSED OVER THE LESION WITH NO DIFFICULTIES FOR STRAIGHTENING OF THE VESSEL. PRE-DILATATION OF THE LESION WAS NOT PERFORMED. AN ATTEMPT WAS MADE TO ADVANCE THE 3.5X12 MM XIENCE PRIME STENT DELIVERY SYSTEM; HOWEVER, IT WOULD NOT CROSS THE LESION. FORCE WAS APPLIED, WHICH RESULTED IN A HYPOTUBE SEPARATION, WHICH WAS ABLE TO BE REMOVED FROM THE ANATOMY WITHOUT ISSUE. A NON-ABBOTT STENT WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. NO ADVERSE PATIENT EFFECTS OR CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE WERE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.