FDA Adverse Event Malfunction Summary report: N

CELL-DYN EMERALD

MDR report key: 3181781 · Received June 21, 2013

Report

Report Number
2919069-2013-00050
Event Type
Malfunction
Date Received
June 21, 2013
Date of Event
May 17, 2013
Report Date
June 5, 2013
Manufacturer
ABBOTT DIAGNOSTICS DIVISION
Product Code
GKZ
PMA / PMN Number
K081495
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A FOLLOWUP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Additional Manufacturer Narrative · 1

THE INVESTIGATION CONSISTED OF REVIEWS OF PRODUCT HISTORY AND PRODUCT LABELING, CONSULTATION WITH THE SUBJECT MATTER EXPERTS, AND INSPECTION OF THE RETURNED CPU BOARD. THE INVESTIGATION CONCLUDED THAT THE IC U200 AND TRANSIENT VOLTAGE SUPPRESSOR D200 WERE BURNED. THESE COMPONENTS WERE BURNT OUT DUE TO A HIGH SURGE OR HIGH VOLTAGE ON THE DC 24V INPUT LINE. POSSIBLE CAUSES COULD BE USAGE OF INCORRECT POWER ADAPTER (DC > 24V) OR EXTERNAL CONDITION LIKE SURGE DUE BY LIGHTNING STORM. THERE ARE NO RISKS OF FIRE ON THE CD EMERALD. NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

THE ABBOTT FIELD SERVICE ENGINEER (FSE) DISCOVERED THE CELL-DYN EMERALD CPU BOARD SHOWED SIGNS OF BEING BURNED. THE FSE WAS REPLACING THE CPU BOARD BECAUSE THE CELL-DYN EMERALD WOULD NOT TURN ON WHEN HE NOTICED THE BURNED CPU BOARD . NO OPERATOR INJURY WAS REPORTED.