FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3181686
·
Received June 14, 2013
Report
- Report Number
- 8020893-2013-01320
- Event Type
- Malfunction
- Date Received
- June 14, 2013
- Date of Event
- May 21, 2013
- Report Date
- May 21, 2013
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT AN 840 VENTILATOR HAD A BLANK SCREEN. NO PT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GUI BACKLIGHT INVERTER PCB. THE UNIT PASSED EXTENDED SELF-TESTING.