FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 3181482 · Received June 21, 2013

Report

Report Number
1416980-2013-16100
Event Type
Malfunction
Date Received
June 21, 2013
Date of Event
May 29, 2013
Report Date
May 29, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS SERVICED AT THE CUSTOMER SITE. THE REPORTED CONDITION OF AN F-94 ALARM WAS CONFIRMED. THE CAUSE OF THE REPORTED CONDITION WAS DUE TO DAMAGED MAIN BATTERIES. TO RESOLVE THE ISSUE THE BATTERIES WERE REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT A FLOGARD INFUSION PUMP EXPERIENCED AN F-94 ALARM. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281883 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1