FDA Adverse Event Malfunction Summary report: N

DHS/DCS-SCR Ø12.5 L85 SST

MDR report key: 3181268 · Received June 21, 2013

Report

Report Number
8030965-2013-03746
Event Type
Malfunction
Date Received
June 21, 2013
Date of Event
May 14, 2013
Report Date
May 23, 2013
Manufacturer
SYNTHES GMBH
Product Code
KTT
PMA / PMN Number
K791619
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. THE MEASURABLE DIMENSIONS WERE FOUND TO BE IN COMPLIANCE WITH THE TECHNICAL DRAWINGS AND AO/ASIF SPECIFICATIONS. THE EXAMINATION OF THE RAW-MATERIAL TESTING CERTIFICATES AND THE MANUFACTURING PAPERS SHOWED NO DEVIATIONS REGARDING MATERIAL ANALYSIS, STRENGTH AND STRUCTURAL STABILITY. THE VALUES WERE IN COMPLIANCE WITH AO/ASIF SPECIFICATION AND WITH THE INTERNATIONAL STANDARD. NO PRODUCT FAULT WAS FOUND. THE INVESTIGATION HAS SHOWN THAT THE POSITIONING GROOVE OF THE SCREW HAS BEEN WIDENED UP DUE TO INADEQUATE HANDLING. THE GROOVE IS EXPANDED AND DAMAGED AND THEREFORE THE PLATE DOES NOT PASS THE SLOTTED END OF THE DHS-DCS SCREW. SUBJECT DEVICE WAS NOT IMPLANTED/EXPLANTED. PLACEHOLDER.

Additional Manufacturer Narrative · 1

A REVIEW OF SYNTHES DEVICE HISTORY RECORDS FOR MANUFACTURING REVEALED NO COMPLAINT RELATED ANOMALIES WERE FOUND. PRODUCT DEVELOPMENT EVALUATION NOT YET COMPLETE. THIS DEVICE WAS USED FOR TREATMENT.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: DURING SURGERY ON AN UNKNOWN DATE, THE LAG SCREW WOULD NOT SLIDE OVER THE DHS PLATE. IT WAS NOTED THAT THE CONNECTION TABS ON THE END OF THE SCREW WERE SLIGHTLY DEFORMED. THE SURGEON USED A NEW LAG SCREW AND IT SLID OVER THE PLATE AS EXPECTED. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
282885 DHS/DCS-SCR Ø12.5 L85 SST KTT SYNTHES GMBH 8301708

Patients

Seq Age Sex Outcome Treatment
1 86 YR