LAG SCREWDRIVER GAMMA3 380X110MM
Report
- Report Number
- 0009610622-2013-00356
- Event Type
- Malfunction
- Date Received
- June 21, 2013
- Date of Event
- May 31, 2013
- Report Date
- May 31, 2013
- Manufacturer
- STRYKER TRAUMA KIEL
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
ONCE THE INVESTIGATION HAS BEEN COMPLETED, ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.
EVALUATION SUMMARY: NO DEVIATIONS WERE FOUND IN THE MANUFACTURING DOCUMENTS OR IN THE MATERIAL. THE RETURNED LAG SCREWDRIVER WAS DOCUMENTED AS MEETING SPECIFICATIONS PRIOR TO DISTRIBUTION. THUS, DEVIATIONS IN MATERIAL AND MANUFACTURING CAN BE EXCLUDED. MARKS FROM MATERIAL DISPLACEMENT AT THE VERY TIP OF THE PEGS AND THE BEND ON THE PEGS THEMSELVES INDICATE THAT THE SCREWDRIVER WAS OPERATED UNDER HIGH FORCE IN CCW / UNTIGHTENING DIRECTION, WHILST THE PEGS WERE NOT ENTIRELY ENGAGED IN THE SLOTS OF THE LAG SCREW. EVALUATION OF ANY DAMAGE CHARACTERISTICS SUGGESTS THAT THE ITEM WAS NOT USED AS INTENDED. IT CANNOT BE EXCLUDED THAT THE ITEM WAS DAMAGED DURING PREVIOUS SURGERIES, BUT THE DAMAGE WOULD THEN HAVE BEEN DETECTED DURING INSPECTION PRIOR TO SURGERY. EVALUATION REVEALED EVIDENCE THAT THE EVENT IS NOT LINKED TO A DEFICIENCY OF THE DEVICE BUT IS RELATED TO IMPROPER USE.
THE CUSTOMER REPORTED TO THE SALES REP, THAT THE SCREWDRIVER DOESN¿T FIT VERY WELL WITH THE COLLUM SCREW DURING THE GAMMA3 PROCEDURE. SALES REP REPORTED THAT ANOTHER TRAY WAS USED TO COMPLETE THE CASE.
THE CUSTOMER REPORTED TO THE SALES REP, THAT THE SCREWDRIVER DOESN¿T FIT VERY WELL WITH THE COLLUM SCREW DURING THE GAMMA3 PROCEDURE. SALES REP REPORTED THAT ANOTHER TRAY WAS USED TO COMPLETE THE CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 281863 | LAG SCREWDRIVER GAMMA3 380X110MM | INSTRUMENT | LXH | STRYKER TRAUMA KIEL | K396526 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |