FDA Adverse Event Malfunction Summary report: N

MINICAP TRANSFER SET

MDR report key: 3181135 · Received June 21, 2013

Report

Report Number
1416980-2013-16078
Event Type
Malfunction
Date Received
June 21, 2013
Date of Event
May 28, 2013
Report Date
May 28, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). ONE USED SET WITH CAP ON DARK BLUE CONNECTOR WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION, LEAK TESTING, CLEAR PASSAGE TEST AND CLAMP FUNCTION TESTS WERE PERFORMED WITH NO ISSUES NOTED. SAMPLE EVALUATION DID NOT CONFIRM THE REPORTED ISSUE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE MINICAP TRANSFER SET HAD LEAKED. THIS OCCURRED AT THE SITE WHERE THE SPIKE IS FOUND, EVEN AFTER THE TWIST CLAMP WAS CLOSED. THERE WAS NO REPORT OF ADVERSE EVENT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281762 MINICAP TRANSFER SET SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1