ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-08692
- Event Type
- Malfunction
- Date Received
- June 20, 2013
- Report Date
- May 23, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/24/2013 WITH THE FOLLOWING FINDINGS: DURING TESTING, THE PUMP POWERED ON WITH A BLANK SCREEN AND NO VIBRATION; THE PUMP WAS EMITTING AUDIBLE TONES. THERE WAS EVIDENT MOISTURE BEHIND THE DISPLAY LENS. THE PUMP FAILED A LEAK TEST DUE TO A CRACK IN THE PUMP¿S CASING. THE PUMP COVER WAS REMOVED AND THERE EVIDENCE OF MOISTURE DAMAGE INTERNALLY. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013, THE PATIENT REPORTED THAT THE PUMP WOULD NOT POWER ON. IT WAS SAID THAT THE PUMP EMITTED A REPLACE BATTERY ALARM PRIOR TO SHUTTING OFF. THE PATIENT DENIED DAMAGE TO THE PUMP CASING OR KEYPAD. IT WAS FURTHER REPORTED THAT THE BATTERY COMPARTMENT WAS DRY AND NOT CORRODED; THE BATTERY CAP WAS SECURE WITHOUT VISIBLE O-RING. NO ADDITIONAL INFORMATION COULD BE OBTAINED FROM THE PUMP AS IT COULD NOT BE TURNED ON. NO ADVERSE EVENT WAS REPORTED RELATED TO THE COMPLAINT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE POWER ISSUE COULD NOT BE RESOLVED DURING TROUBLE SHOOTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 279819 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR |