FDA Adverse Event Malfunction Summary report: N

PARADIGM INSULIN INFUSION PUMP

MDR report key: 3180937 · Received June 20, 2013

Report

Report Number
3004209178-2013-94861
Event Type
Malfunction
Date Received
June 20, 2013
Date of Event
June 17, 2013
Report Date
June 17, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
LZG
PMA / PMN Number
K040676
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED DURING BASIC OCCLUSION TEST AND UNABLE TO PRIME DURING PRIME TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP ALARMED, AND THE DRIVE SUPPORT CAP WAS LOOSE. ADVISED THE CALLER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281371 PARADIGM INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-715WWS

Patients

Seq Age Sex Outcome Treatment
1