FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3180919 · Received June 20, 2013

Report

Report Number
3004209178-2013-94885
Event Type
Injury
Date Received
June 20, 2013
Date of Event
May 23, 2013
Report Date
June 3, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED BEING HOSPITALIZED DUE TO LOW BLOOD GLUCOSE OF 40MG/DL. THE CUSTOMER STATED THAT HE FEELS THE INSULIN PUMP DELIVERED INSULIN WHILE SLEEPING. TROUBLESHOOTING WAS PERFORMED, AND THE DRIVE SUPPORT CAP APPEARS TO BE NORMAL. REVIEWED THE PROGRAMMING IN THE DEVICE, AND THE BOLUS HISTORY REVEALED TWO BOLUSES OF 9.0 UNITS EACH PRIOR TO THE PARAMEDIC'S ARRIVAL. THE CUSTOMER MENTIONED THAT HE WAS INCOHERENT AT THAT TIME. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
279709 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 30 YR Hospitalization