PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2013-94897
- Event Type
- Injury
- Date Received
- June 20, 2013
- Date of Event
- May 31, 2013
- Report Date
- May 31, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A FAULTY FORCE SENSOR. FURTHER TESTING COULD NOT BE PERFORMED DUE TO THE PRIME ANOMALY. THE DEVICE WAS RECEIVED WITH SCRATCHED DISPLAY WINDOW AND STRIPPED BATTERY CAP COIN SLOT. THE DRIVE SUPPORT DISK WAS INSPECTED AND NO ANOMALY NOTED. A TEST RESERVOIR WAS INSTALLED AND FITTED PROPERLY. NO CONNECTION ANOMALIES FOUND AT THE RESERVOIR TUBE.
IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO DIABETES KETOACIDOSIS AND HIGH BLOOD GLUCOSE OF 754MG/DL. THE CUSTOMER EXPERIENCED NAUSEA AND VOMITING. TROUBLESHOOTING WAS PERFORMED, AND THE DRIVE SUPPORT CAP APPEARS TO BE KICKED UP A HAIR ON ONE SIDE. THE TIME, DATE, BASAL RATES, AND BOLUS WIZARD SETTINGS WERE CORRECT. REVIEWED THE ALARM HISTORY AND FOUND MULTIPLE NO DELIVERY ALARMS. ASSISTED THE CALLER TO RUN A MANUAL PRIME TEST AND THE INSULIN DID EXIT. PERFORMED A HIGH PRESSURE TEST AND PASSED. ADVISED THE FATHER THAT THE INSULIN PUMP IS FUNCTIONING AS DESIGNED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 281258 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 YR | Hospitalization |