FDA Adverse Event Malfunction Summary report: N

COULTER® LH 500 HEMATOLOGY ANALYZER

MDR report key: 3180904 · Received June 20, 2013

Report

Report Number
1061932-2013-01233
Event Type
Malfunction
Date Received
June 20, 2013
Date of Event
May 21, 2013
Report Date
May 21, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K042724
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) OPENED THE RIGHT SIDE OF THE INSTRUMENT PANEL AND DISCOVERED A HOLE IN THE BLACK I-BEAM TUBING WHICH CAUSED THE FLUID LEAK. THE FSE REPLACED THE TUBING AND PERFORMED MULTIPLE TESTS SUCCESSFULLY. SERVICE ACTIVITY WAS VERIFIED TO MEET SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURE. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED PERFORMANCE SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED CLENZ CLEANER SOLUTION LEAKED ONTO THE COUNTER WHILE PERFORMING SYSTEM STARTUP INVOLVING THE COULTER LH 500 HEMATOLOGY ANALYZER. THE OPERATOR WAS WEARING PROPER PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND EYE PROTECTION AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. PATIENT RESULTS WERE NOT IMPACTED. THERE WAS NO PATIENT CONSEQUENCE ASSOCIATED WITH THIS EVENT. THE CUSTOMER DISCONTINUED USE OF THE INSTRUMENT AND A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT. THE FACILITY HAS AN EXPOSURE CONTROL OR RISK MANAGEMENT PLAN IN PLACE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
279653 COULTER® LH 500 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1