FDA Adverse Event Malfunction Summary report: N

COULTER® LH 750 HEMATOLOGY ANALYZER

MDR report key: 3180859 · Received June 20, 2013

Report

Report Number
1061932-2013-01226
Event Type
Malfunction
Date Received
June 20, 2013
Date of Event
May 25, 2013
Report Date
May 25, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K061574
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2013 TO EVALUATE THE INSTRUMENT. THE FSE FOUND THAT THE LYSE SAMPLE LINE THROUGH THE PINCH VALVE VL65 HAD POPPED OFF THE FITTING AT THE DIFFERENTIAL MIXING CHAMBER. THE FSE REPLACED THE TUBING AND THE INSTRUMENT RAN WITHOUT ANY LEAKS. THE REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES. PER LABELING, BECKMAN COULTER URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IS NOT LIMITED TO PROTECTIVE EYEWEAR, GLOVES AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER (BEC) REPORTING A LEAK INSIDE THE COULTER® LH 750 HEMATOLOGY ANALYZER. THE VOLUME OF THE LEAK WAS ABOUT 3 ML AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND PROTECTIVE EYEWEAR AT THE TIME OF THE INCIDENT. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED. THE LABORATORY'S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. MEDICAL ATTENTION WAS NOT SOUGHT. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION TO THE REPORTED EVENT.