FDA Adverse Event Malfunction Summary report: N

STERRAD 200 STERILIZATION SYSTEM

MDR report key: 3180855 · Received June 20, 2013

Report

Report Number
2084725-2013-00273
Event Type
Malfunction
Date Received
June 20, 2013
Date of Event
June 11, 2012
Report Date
June 11, 2013
Manufacturer
ADVANCED STERILIZATION PRODUCTS
Product Code
MLR
PMA / PMN Number
K030429
Removal / Correction Number
Z-1026-1027-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ROOT CAUSE IDENTIFIED IN CAPA WAS FOUND TO BE PREMATURE FAILURE/SATURATED OIL MIST FILTER OR VACUUM PUMP OIL ASP INVESTIGATION SUMMARY: THE INVESTIGATION INCLUDED A REVIEW OF THE DEVICE HISTORY RECORD, SERVICE HISTORY, TRENDING OF THE PRODUCT MALFUNCTION CODE, FAILURE MODE AND EFFECTS ANALYSIS, HEALTH HAZARD EVALUATION, SYSTEM HAZARD AND USER MISUSE ANALYSIS, AND CAPA. THE DHR (DEVICE HISTORY RECORD) WAS REVIEWED AND NO ISSUES RELATING TO THE FAILURE MODE WERE NOTED. THE INVOLVED UNIT MET MANUFACTURER SPECIFICATIONS AT THE TIME OF RELEASE. THE SERVICE HISTORY FOR THIS UNIT FOR THE PAST 6 MONTHS (DECEMBER 2012 TO JUNE 2013) DID NOT REVEAL A SIGNIFICANT TREND FOR THIS SAME ISSUE. THE TREND OF THE PRODUCT MALFUNCTION CODE ODOR/SMELLS WAS COMPLETED FROM JANUARY 2013 THROUGH DECEMBER 2013 AND REVEALED A SIGNIFICANT TREND WHICH WAS ADDRESSED THROUGH CAPA. THE FMEA (FAILURE MODE AND EFFECTS ANALYSIS) REVEALED THE RISK PRIORITY NUMBER FOR THIS FAILURE MODE IS 128 WHICH IS GREATER THAN THE ACCEPTABLE LIMIT. A CAPA WAS OPENED TO ADDRESS THIS ISSUE. THE HHE (HEALTH HAZARD EVALUATION) WAS REVIEWED FOR THE RISK OF ODOR AND SMELL EXPOSURE. THE SEVERITY AND OCCURRENCE FOR THE GENERAL POPULATION WERE LIMITED (TRANSIENT, MINOR IMPAIRMENT, NO MEDICAL TREATMENT REQUIRED) AND THE PRODUCT PROBLEM HAS BEEN KNOWN TO RESULT IN THE IDENTIFIED HARM, BUT ONLY OCCASIONALLY AND/OR UNDER UNUSUAL CIRCUMSTANCES. THE SHUMA (SYSTEM HAZARD AND USER MISUSE ANALYSIS) DETERMINED THE RISK IS AS LOW AS REASONABLY PRACTICABLE FOR EXPOSURE TO ODOR OR ODORANTS. THE CAPA (CORRECTIVE AND PREVENTATIVE ACTION) IDENTIFIED THE ROOT CAUSE FOR THE ODOR/SMELLS ISSUE AS: PREMATURE FAILURE OF THE USED AND SATURATED STERRAD® 200 OIL MIST FILTER CAUSED OIL VAPOR EMISSIONS THAT EXACERBATED THE ODOR/SMELL COMPLAINTS REPORTED FOR THE STERRAD® 200 SYSTEM. THE USE OF A LESS OXIDATIVELY STABLE VACUUM PUMP OIL CAUSED THE ODOR/SMELLS FOR THE STERRAD® 200 SYSTEM. THE STERILIZER WAS NOT SERVICED AS A RESULT OF THIS ISSUE; HOWEVER, A PLANNED MAINTENANCE WAS LATER PERFORMED ON (B)(6) 2013. NO ODOR WAS DETECTED AFTER THIS SERVICE AND THE STERILIZER WAS LEFT IN WORKING ORDER. NO PARTS WERE RETURNED FOR FURTHER EVALUATION. ASP WILL CONTINUE TO TRACK AND TREND THIS ISSUE.

Description of Event or Problem · 1

AN INTERNATIONAL CUSTOMER REPORTED AN EVENT OF A "VINEGAR ODOR" EMITTING FROM THE STERRAD 200 STERILIZER. THERE WAS NO REPORT OF ANY HUMAN REACTION. A FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY EVENT DATED (B)(6) 2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
280388 STERRAD 200 STERILIZATION SYSTEM STERRAD EQUIPMENT (MLR) MLR ADVANCED STERILIZATION PRODUCTS NA 0202090028

Patients

Seq Age Sex Outcome Treatment
1