FDA Adverse Event Malfunction Summary report: N

WEBSTER ELECTROPHYSIOLOGY CATHETER WITH AUTO ID

MDR report key: 3180799 · Received June 20, 2013

Report

Report Number
9673241-2013-00205
Event Type
Malfunction
Date Received
June 20, 2013
Date of Event
June 11, 2013
Report Date
June 11, 2013
Manufacturer
BIOSENSE WEBSTER, INC. (JUAREZ)
Product Code
DRF
PMA / PMN Number
K892265
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

MULTIPLE ATTEMPTS HAVE BEEN MADE TO REQUEST FOR THE COMPLAINT PRODUCT TO BE RETURNED FOR ANALYSIS. PRODUCT WAS NOT RETURNED FOR INVESTIGATION AS INITIALLY REPORTED. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED. THE DHR REVIEW VERIFIES THAT THE DEVICES WERE MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. MANUFACTURER'S REF. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED DURING AN ATRIOVENTRICULAR REENTRANT TACHYCARDIA (AVRT)/(WPW) PROCEDURE, THIS WEBSTER 4 POLE W/AUTO ID CATHETER CAUSED GLOBAL NOISE ON THE CARTO 3 SYSTEM AS WELL AS THE RECORDING SYSTEM. CHANGING THE CATHETER CABLE DID NOT RESOLVE THE ISSUE. CHANGING THIS CATHETER RESOLVED THE ISSUE. THE PROCEDURE WAS CONTINUED WITH NO PATIENT INJURY REPORTED. UPON REQUEST, THE BWI FIELD REPRESENTATIVE PROVIDED ADDITIONAL INFORMATION ON THE EVENT. THE PHYSICIAN WAS NOT ABLE TO INTERPRET THE SIGNALS BECAUSE OF THE PRESENCE OF THE NOISE. THE NOISE WAS ON ALL THE CHANNELS, INCLUDING THE 12 LEADS OF BODY SURFACE (BS) ECG¿S AND INTRACARDIAC (IC) SIGNALS ON BOTH THE CARTO 3 SYSTEM AND THE EP RECORDING SYSTEM AT THE SAME TIME. THE ISSUE OCCURRED WHEN THEY WERE CONNECTING THE CATHETER. THE SEVERITY OF THE NOISE ON ALL THE CHANNELS ON BOTH THE CARTO 3 SYSTEM AND THE RECORDING SYSTEM AT THE SAME TIME IS INDICATIVE OF A REPORTABLE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
279586 WEBSTER ELECTROPHYSIOLOGY CATHETER WITH AUTO ID CATHETER, ELECTRODE RECORDING DRF BIOSENSE WEBSTER, INC. (JUAREZ) D-1079-259-S 14008145

Patients

Seq Age Sex Outcome Treatment
1