FDA Adverse Event Malfunction Summary report: N

COULTER® LH 500 HEMATOLOGY ANALYZER

MDR report key: 3180791 · Received June 20, 2013

Report

Report Number
1061932-2013-01127
Event Type
Malfunction
Date Received
June 20, 2013
Date of Event
May 27, 2013
Report Date
May 27, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K032000
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE FSE OBSERVED THAT THE INSTRUMENT WAS LEAKING CLEANER. THE FSE FOUND THE SOURCE OF THE LEAK WAS THE TUBING AT PINCH VALVE (PV 44). THE PV44 CONTROLS THE PRESSURIZED DILUENT FOR FLUSHING THE UPPER PART OF THE FLOWCELL. THE FSE REPLACED THE AFFECTED TUBING AND ALSO REPLACED ALL DILUTER TUBING AND PERFORMED PREVENTIVE MAINTENANCE (PM) AS IT WAS COMING DUE. AFTER SERVICING THE INSTRUMENT, NO FURTHER EVIDENCE OF LEAKING WAS OBSERVED. THE CAUSE OF THE LEAK WAS THE TUBING AT PV44. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO BECKMAN COULTER (BEC) A LEAK COMING FROM THE RIGHT SIDE OF THE COULTER LH 500 HEMATOLOGY ANALYZER WHILE RUNNING A STARTUP. THE VOLUME OF THE LEAK IS UNKNOWN AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. IT COULD NOT BE DETERMINED IF THE INSTRUMENT GENERATED ANY ERROR MESSAGES OR FLAGS AS A RESULT OF THIS EVENT. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED BY THE CUSTOMER. THE LABORATORY¿S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE), CONSISTING OF GLOVES, EYE PROTECTION, AND A LABORATORY COAT AT THE TIME OF THE OCCURRENCE. NO ONE CAME IN CONTACT WITH THE FLUID. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. MEDICAL ATTENTION WAS NOT SOUGHT. NO ERRONEOUS RESULTS WERE GENERATED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281496 COULTER® LH 500 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER LH 500 N/A

Patients

Seq Age Sex Outcome Treatment
1