FDA Adverse Event Malfunction Summary report: N

COULTER® LH 750 HEMATOLOGY ANALYXZER

MDR report key: 3180784 · Received June 20, 2013

Report

Report Number
1061932-2013-01130
Event Type
Malfunction
Date Received
June 20, 2013
Date of Event
May 28, 2013
Report Date
May 28, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K011342
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE ON (B)(4) 2013. THE FSE INSPECTED THE INSTRUMENT AND DISCOVERED THAT THE TUBING FROM THE NEEDLE TO FLUID DETECTOR FD1 WAS LOOSE AT THE FITTING OF THE FLUID DETECTOR AND WAS LEAKING. THE FSE REPLACED THE TUBING AND THE INSTRUMENT RAN WITHOUT ANY LEAKS. THE REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES. THE CAUSE OF THE LEAK WAS THAT THE TUBING FROM THE NEEDLE TO FLUID DETECTOR FD1 WAS LOOSE AT THE FITTING OF THE FLUID DETECTOR. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO BECKMAN COULTER (BEC) A LEAK AT THE BLOOD SAMPLING VALVE (BSV) OF THE COULTER LH 750 HEMATOLOGY ANALYZER. THE VOLUME OF THE LEAK WAS ABOUT 3 ML AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE LEAK DID NOT IMPACT ELECTRICAL OR OPTICAL PERFORMANCE OF SYSTEM. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED BY THE CUSTOMER. THE LABORATORY¿S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE), CONSISTING OF GLOVES, EYE GLASSES AND A LABORATORY COAT AT THE TIME OF THE OCCURRENCE. NO ONE CAME IN CONTACT WITH THE FLUID. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. MEDICAL ATTENTION WAS NOT SOUGHT. NO DISCREPANT PATIENT RESULTS WERE GENERATED IN CONNECTION WITH THIS EVENT. THERE WAS NO DEATH, INJURY OR CHANGE TO PATIENT TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
279573 COULTER® LH 750 HEMATOLOGY ANALYXZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER LH 750 N/A

Patients

Seq Age Sex Outcome Treatment
1