FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3180728 · Received June 20, 2013

Report

Report Number
2531779-2013-08673
Event Type
Malfunction
Date Received
June 20, 2013
Report Date
May 27, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

ON EXAMINATION, THE KEYPAD WAS INTACT WITHOUT DAMAGE. ON TESTING, NONE OF THE KEYPAD BUTTONS RESPONDED TO PRESSES EXCEPT THE CONTRAST BUTTON. ON TESTING, THE PUMP POWERED ON WITH FUNCTIONAL AUDITORY AND VIBRATORY ALARMS. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED NO DAMAGE, DEFECTS OR CONTAMINATION OF THE COMPONENTS UNDER THE KEYPAD COVER. THE BOLUS BUTTON COVER WAS REMOVED FOR INVESTIGATION AND REVEALED THE BUTTON WAS STUCK IN THE ¿ON ¿POSITION.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. IT WAS REPORTED THAT THE UP ARROW AND DOWN ARROW KEYPAD BUTTONS WERE UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
280654 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 13 YR