ROTALINK? PLUS
Report
- Report Number
- 2134265-2013-04290
- Event Type
- Injury
- Date Received
- June 20, 2013
- Date of Event
- May 22, 2013
- Report Date
- May 23, 2013
- Manufacturer
- BOSTON SCIENTIFIC - CORK
- Product Code
- MCX
- PMA / PMN Number
- P900056
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4).
DEVICE EVALUATED BY MANUFACTURER: THE BURR WAS MICROSCOPICALLY EXAMINED. THE ANNULUS WAS FOUND TO BE FLARED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT. (B)(4).
SAME CASE AS MDR ID 2134265-2013-04289, 2134265-2013-04291. IT WAS REPORTED THAT DURING PERCUTANEOUS CORONARY INTERVENTION, THE WIRE WAS DETACHED/SEPARATED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE PROXIMAL (LAD) LEFT ANTERIOR DESCENDING. A ROTAWIRE EX SUPPORT GUIDEWIRE WAS SELECTED TO GUIDE THE BURR. IT WAS NOTED THAT ABLATION WAS PERFORMED AT LEFT CIRCUMFLEX ARTERY AND POST DILATATION WAS PERFORMED USING A NON-BSC BALLOON CATHETER AND THEN A PROMUS ELEMENT PLUS STENT 3.0 X 20 WAS DEPLOYED. THEN THE ROTAWIRE WAS ADVANCED BETWEEN THE LAD AND AT THE FIRST DIAGONAL BRANCH, AND ABLATION WAS PERFORMED AT OSTIAL PART OF LAD. WHEN ATTEMPTING TO PERFORM 3RD ABLATION THE ROTAWIRE WAS DETACHED. THE DETACHED WIRE WAS REMOVED USING A NON-BSC DEVICE. AFTER REMOVING THE DETACHED WIRE ANOTHER ABLATION WAS PERFORMED USING A ROTABURR 1.5 AND ANOTHER ROTAWIRE EX SUPPORT GUIDEWIRE. UPON 4TH ABLATION, IT WAS THEN NOTED THAT ANOTHER ROTAWIRE WAS DETACHED AT FIRST DIAGONAL BRANCH. THE DETACHED WIRE WAS NOT REMOVED FROM THE PATIENT¿S BODY. POST DILATATION WAS PERFORMED USING A NON-BSC PRODUCT AT THE OSTIAL PART OF THE FIRST DIAGONAL BRANCH THEN ABLATION WAS DISCONTINUED. SAME PROCEDURE WAS PERFORMED TO LAD. POST DILATATION WAS PERFORMED AT THE MIDDLE AND PROXIMAL LAD USING A DIFFERENT NON-BSC BALLOON CATHETER. A PROMUS ELEMENT PLUS STENT 2.5 X 15MM WAS THEN DEPLOYED AT THE MIDDLE LAD WITH A NON-BSC BALLOON CATHETER. A NON-BSC STENT WAS DEPLOYED AT PROXIMAL LAD WITH A NON-BSC BALLOON CATHETER. THEN A KISSING BALLOON TECHNIQUE WAS PERFORMED FOR THE TREATMENT OF THE PROXIMAL LAD AND AT THE FIRST DIAGONAL BRANCH. HUGGING BALLOON TECHNIQUE WAS PERFORMED AND ADDITIONAL DILATATION WAS MADE AT THE PROXIMAL LAD WITH A NON-BSC BALLOON CATHETER. PROCEDURE WAS THEN COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT¿S CONDITION WAS GOOD.
SAME CASE AS MDR ID 2134265-2013-04289, 2134265-2013-04291. IT WAS REPORTED THAT DURING PERCUTANEOUS CORONARY INTERVENTION, THE WIRE WAS DETACHED/SEPARATED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE PROXIMAL (LAD) LEFT ANTERIOR DESCENDING. A ROTAWIRE EX SUPPORT GUIDEWIRE WAS SELECTED TO GUIDE THE BURR. IT WAS NOTED THAT ABLATION WAS PERFORMED AT LEFT CIRCUMFLEX ARTERY AND POST DILATATION WAS PERFORMED USING A NON-BSC BALLOON CATHETER AND THEN A PROMUS ELEMENT PLUS STENT 3.0 X 20 WAS DEPLOYED. THEN THE ROTAWIRE WAS ADVANCED BETWEEN THE LAD AND AT THE FIRST DIAGONAL BRANCH, AND ABLATION WAS PERFORMED AT OSTIAL PART OF LAD. WHEN ATTEMPTING TO PERFORM 3RD ABLATION THE ROTAWIRE WAS DETACHED. THE DETACHED WIRE WAS REMOVED USING A NON-BSC DEVICE. AFTER REMOVING THE DETACHED WIRE ANOTHER ABLATION WAS PERFORMED USING A ROTABURR 1.5 AND ANOTHER ROTAWIRE EX SUPPORT GUIDEWIRE. UPON 4TH ABLATION, IT WAS THEN NOTED THAT ANOTHER ROTAWIRE WAS DETACHED AT FIRST DIAGONAL BRANCH. THE DETACHED WIRE WAS NOT REMOVED FROM THE PATIENT¿S BODY. POST DILATATION WAS PERFORMED USING A NON-BSC PRODUCT AT THE OSTIAL PART OF THE FIRST DIAGONAL BRANCH THEN ABLATION WAS DISCONTINUED. SAME PROCEDURE WAS PERFORMED TO LAD. POST DILATATION WAS PERFORMED AT THE MIDDLE AND PROXIMAL LAD USING A DIFFERENT NON-BSC BALLOON CATHETER. A PROMUS ELEMENT PLUS STENT 2.5 X 15MM WAS THEN DEPLOYED AT THE MIDDLE LAD WITH A NON-BSC BALLOON CATHETER. A NON-BSC STENT WAS DEPLOYED AT PROXIMAL LAD WITH A NON-BSC BALLOON CATHETER. THEN A KISSING BALLOON TECHNIQUE WAS PERFORMED FOR THE TREATMENT OF THE PROXIMAL LAD AND AT THE FIRST DIAGONAL BRANCH. HUGGING BALLOON TECHNIQUE WAS PERFORMED AND ADDITIONAL DILATATION WAS MADE AT THE PROXIMAL LAD WITH A NON-BSC BALLOON CATHETER. PROCEDURE WAS THEN COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT¿S CONDITION WAS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 280829 | ROTALINK? PLUS | CATHETER, CORONARY, ATHERECTOMY | MCX | BOSTON SCIENTIFIC - CORK | H749236310030 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | GUIDEWIRE: SION BLUE| STENT: PROMUSELMNET 2.5X28/3.0X20| BALLOON CATHETER: NC TREK| GUIDING CATHETER: LAUNCHER LAUNCHER / EBU3.5SH |