FDA Adverse Event Malfunction Summary report: N

LIGACLIP** ENDOSCOPIC CLIP APPLIER

MDR report key: 3180419 · Received June 20, 2013

Report

Report Number
3005075853-2013-03156
Event Type
Malfunction
Date Received
June 20, 2013
Report Date
June 11, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS OF THE DEVICE FOUND THAT IT WAS RECEIVED WITH THE JAWS MISALIGNED. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON FIRING OF THE DEVICE, THE CLIPS WERE FED AS INTENDED; HOWEVER, DUE TO THE MISALIGNED CONDITION OF THE JAWS SCISSOR CLIPS WERE FORMED. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. NO INCIDENT RELATED TO THE REPORTED EVENT WAS OBSERVED DURING THE BATCH RECORD REVIEW.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE WAS FIRING SCISSORED CLIPS. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT IMPACT REPORTED. THESE WERE ALL THE DETAIL PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281106 LIGACLIP** ENDOSCOPIC CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA K4CL94

Patients

Seq Age Sex Outcome Treatment
1