FDA Adverse Event Malfunction Summary report: N

INSTINCT ENDOSCOPIC HEMOCLIP

MDR report key: 3180263 · Received June 13, 2013

Report

Report Number
1037905-2013-00386
Event Type
Malfunction
Date Received
June 13, 2013
Date of Event
May 10, 2013
Report Date
May 13, 2013
Manufacturer
WILSON-COOK MEDICAL INC.
Product Code
MND
PMA / PMN Number
K121505
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSIONS: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVALUATION. PRIOR TO DISTRIBUTION, ALL INSTINCT ENDOSCOPIC HEMOCLIPS ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MANUFACTURED PRIOR TO IMPLEMENTATION OF THIS CORRECTIVE ACTION. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPALINT TRENDS.

Description of Event or Problem · 1

DURING AN ENDOSCOPIC PROCEDURE, A COOK INSTINCT ENDOSCOPIC HEMOCLIP WAS USED. THE CLIP DEPLOYED, BUT THE HOOK AT THE DISTAL END OF THE DRIVE WIRE WOULD NOT RETRACT BACK INTO THE ENDOSCOPE. THIS REASONABLY SUGGESTS THAT THE DRIVE WIRE DISCONNECTED FROM THE HANDLE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PT'S BODY. THE PT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
269238 INSTINCT ENDOSCOPIC HEMOCLIP MND, LIGATOR, ESOPHAGEAL MND WILSON-COOK MEDICAL INC. W3261779

Patients

Seq Age Sex Outcome Treatment
1 ENDOSCOPE (UNKNOWN TYPE)