INSTINCT ENDOSCOPIC HEMOCLIP
Report
- Report Number
- 1037905-2013-00386
- Event Type
- Malfunction
- Date Received
- June 13, 2013
- Date of Event
- May 10, 2013
- Report Date
- May 13, 2013
- Manufacturer
- WILSON-COOK MEDICAL INC.
- Product Code
- MND
- PMA / PMN Number
- K121505
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- UNKNOWN
Narratives
INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSIONS: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVALUATION. PRIOR TO DISTRIBUTION, ALL INSTINCT ENDOSCOPIC HEMOCLIPS ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MANUFACTURED PRIOR TO IMPLEMENTATION OF THIS CORRECTIVE ACTION. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPALINT TRENDS.
DURING AN ENDOSCOPIC PROCEDURE, A COOK INSTINCT ENDOSCOPIC HEMOCLIP WAS USED. THE CLIP DEPLOYED, BUT THE HOOK AT THE DISTAL END OF THE DRIVE WIRE WOULD NOT RETRACT BACK INTO THE ENDOSCOPE. THIS REASONABLY SUGGESTS THAT THE DRIVE WIRE DISCONNECTED FROM THE HANDLE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PT'S BODY. THE PT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 269238 | INSTINCT ENDOSCOPIC HEMOCLIP | MND, LIGATOR, ESOPHAGEAL | MND | WILSON-COOK MEDICAL INC. | W3261779 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ENDOSCOPE (UNKNOWN TYPE) |