SITTER SELECT
Report
- Report Number
- 2020362-2013-00246
- Event Type
- Malfunction
- Date Received
- June 4, 2013
- Report Date
- May 6, 2013
- Manufacturer
- J. T. POSEY CO.
- Product Code
- KMI
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- OTHER
Narratives
RESULTS: EVAL OF THE RETURNED PRODUCT DID NOT CONFIRM THE REPORTED ISSUE; THE ALARM WAS TESTED THREE TIMES AT FIVE MIN INTERVALS WITH NEW BATTERIES AND THE ALARM POWERED UP EACH TIME WITHOUT ANY INTERMITTENCY OBSERVED. HOWEVER, THE ALARM DOES NOT SOUND WHEN THE SENSOR IS CONNECTED AND WEIGHT IS OFF THE SENSOR, WHEN THE SENSOR IS UNPLUGGED OR WHEN THE MAGNET IS REMOVED. THE RED LED ON THE NURSE CALL FIXTURE LIGHTS WHEN THE NURSE CALL FUNCTION IS TESTED. NO OTHER FUNCTIONAL TESTS CAN BE PERFORMED SINCE THE ALARM DOES NOT SOUND AT ALL. THE ALARM WAS RECEIVED WITH BATTERIES INSIDE THE UNIT AND IS VISIBLY LEAKING INSIDE THE BATTERY COMPARTMENT BUT THE BATTERY CONTACTS DO NOT HAVE CORROSION ON THEM. THE BATTERY SPRINGS ARE SLIGHTLY BENT. DELAY SWITCH IS AT 4 SECONDS. AQUALERT WATER INDICATOR LABEL IS MISSING. NOTE: INSTRUCTIONS FOR USE: DO NOT USE IF; BATTERY DOOR IS MISSING, BATTERY DOOR IS DAMAGED; ALARM CASE IS DAMAGED; OR ALARM CASE IS CRACKED. WHEN ACCESSING THE BATTERY COMPARTMENT SLIDE BATTERY DOOR OPEN AND FLIP IT UP. DO NOT ATTEMPT TO REMOVE BATTERY DOOR; IT DOES NOT SEPARATE FROM ALARM. CAREFULLY REMOVE BATTERIES TO AVOID DAMAGE TO BATTERY DOOR. INSERT FOUR (4) NEW "AA" ALKALINE BATTERIES. DO NOT MIX OLD AND NEW BATTERIES, OR BATTERY BRANDS WITHIN A BATTERY PACK (4 BATTERIES). USE OF MIXED BATTERIES OR BATTERIES INSTALLED INCORRECTLY MAY CAUSE BATTERY DAMAGE, AND MAY DAMAGE THE ALARM. REMOVE ANY ALARM FROM USE AND SEND TO THE APPROPRIATE FACILITY AUTHORITY IF BATTERIES ARE DAMAGED OR CORRODED OR THE BATTERY COMPARTMENT HAS SIGNS OF PREVIOUS BATTERY CORROSION SUCH AS WHITE POWDER RESIDUE. DO NOT USE THE ALARM IF BATTERY DAMAGE HAS BEEN DETECTED. (B)(4).
THE CUSTOMER REPORTED THE ALARM WILL NOT POWER ON. THE BATTERIES HAVE BEEN REPLACED WITH A NEW SUPPLY. THE CUSTOMER REPORTED THERE IS NO PHYSICAL DAMAGE TO THE OUTSIDE OF THE ALARM. THE CUSTOMER DID NOT PROVIDE A DATE WHEN THIS WAS DISCOVERED. NO PT INCIDENT OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 247598 | SITTER SELECT | KMI | J. T. POSEY CO. | 8361 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |