FDA Adverse Event
Malfunction
Summary report: N
SM304 M-SERIES W/ZOOM
MDR report key: 3180002
·
Received June 20, 2013
Report
- Report Number
- 0001831750-2013-05598
- Event Type
- Malfunction
- Date Received
- June 20, 2013
- Date of Event
- April 29, 2013
- Report Date
- April 29, 2013
- Manufacturer
- STRYKER MEDICAL-KALAMAZOO
- Product Code
- INK
- PMA / PMN Number
- K022309
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PREVIOUSLY, THE MDR INITIAL WAS ISSUED WITHOUT THE PMA/510(K)#. THIS HAS BEEN INCLUDED IN THIS MDR FOLLOW-UP REPORT.
Description of Event or Problem · 1
IT WAS REPORTED VIA REPAIR WORK ORDER THAT THE ZOOM DRIVE WAS INTERMITTENT DUE TO DAMAGED LOAD CELL WELDMENT. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Description of Event or Problem · 1
IT WAS REPORTED VIA REPAIR WORK ORDER THAT THE ZOOM DRIVE WAS INTERMITTENT DUE TO DAMAGED LOAD CELL WELDMENT. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 279671 | SM304 M-SERIES W/ZOOM | STRETCHER, WHEELED, POWERED | INK | STRYKER MEDICAL-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |