FDA Adverse Event Malfunction Summary report: N

SABRATEK 6060 PUMP

MDR report key: 317974 · Received February 26, 2001

Report

Report Number
MW1021196
Event Type
Malfunction
Date Received
February 26, 2001
Date of Event
February 21, 2001
Report Date
February 26, 2001
Manufacturer
BAXTER HEALTHCARE
Product Code
FRN
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
NV, US
Reporter Occupation
PHARMACIST

Narratives

Description of Event or Problem · 1

PT CALLED - UNABLE TO GET PUMP TO WORK. DISPLAY READS MALFUNCTION-18. PHARMACY WILL DELIVER REPLACEMENT PUMP AND CHECK CURRENT PUMP FOR MALFUNCTION. UPON OBSERVATION IT WAS NOTED THAT TOTAL PARENTERAL NUTRITION SOLUTION HAD LEAKED INTO THE PERISTALTIC MOTOR UNIT. TUBING WAS PREVIOUSLY REMOVED AND DISCARDED SO IT WAS UNAVAILABLE FOR EXAMINATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
7710 SABRATEK 6060 PUMP AMBULATORY INFUSION PUMP FRN BAXTER HEALTHCARE 6060 *

Patients

Seq Age Sex Outcome Treatment
1 57 YR Other LOT# 001888.| 1) 560112-GEL TUBING FOR SABRATEK 6060 PUMP,