NEOPUFF INFANT RESUSCITATOR
Report
- Report Number
- 9611451-2013-00479
- Event Type
- Malfunction
- Date Received
- June 19, 2013
- Date of Event
- May 21, 2013
- Report Date
- May 21, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTL
- PMA / PMN Number
- K892885
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- NURSE
Narratives
(B)(4). WE ARE IN THE PROCESS OF OBTAINING THE COMPLAINT DEVICES FROM THE HEALTHCARE FACILITY FOR EVALUATION. WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
(B)(4) IT WAS REPORTED THAT THE GAS OUTLET PORT WAS BROKEN ON THREE RD900 NEOPUFF DEVICES, TWO FROM LOT 091207 AND ONE FROM 090203. METHOD: ONLY THE 900RD010 T-PIECE CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) WHERE IT WAS VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED THAT THE T-PIECE CIRCUIT WAS RETURNED WITH A BROKEN GAS OUTLET PORT FROM A NEOPUFF DEVICE STILL CONNECTED TO ONE END OF THE RETURNED T-PIECE CIRCUIT. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOTS 090203 AND 091207. CONCLUSION: THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANTS UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. BEING A PORTABLE DEVICE, THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. THE NEOPUFF TECHNICAL MANUAL WARNS AGAINST DROPPING THE NEOPUFF OR SUBJECTING IT TO IMPACT DAMAGE WHICH MAY CAUSE THE UNIT TO OPERATE INCORRECTLY. IF THE NEOPUFF IS SUSPECTED TO HAVE BEEN DAMAGED, THE MANOMETER AND VALVE SYSTEM SHOULD BE PERFORMANCE TESTED. THE HOSPITAL CORRECTLY IDENTIFIED THE FAULT DURING SET UP AS INDICATED IN OUR UI. THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: "DROPPING THE NEOPUFF / PERIVENT INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT".
A HEALTHCARE FACILITY IN (B)(6) REPORTED THAT THE GAS OUTLET PORT ON THREE RD900 NEOPUFF RESUSCITATORS BROKE WHEN THEY CONNECTED THE 900RD010 T-PIECE CIRCUIT. THIS WAS FOUND DURING SET UP AND PRIOR TO PATIENT USE.
A HEALTHCARE FACILITY IN (B)(6) REPORTED THAT THE GAS OUTLET PORT ON THREE RD900 NEOPUFF RESUSCITATORS BROKE WHEN THEY CONNECTED THE 900RD010 T-PIECE CIRCUIT. THIS WAS FOUND DURING SET UP AND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 278632 | NEOPUFF INFANT RESUSCITATOR | BTL | BTL | FISHER & PAYKEL HEALTHCARE LTD | RD900 | 090203 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 900RD010 T-PIECE CIRCUIT| 900RD010 T-PIECE CIRCUIT |