FDA Adverse Event Malfunction Summary report: N

PFNA-II BLADE L85 TAN

MDR report key: 3174428 · Received June 18, 2013

Report

Report Number
8030965-2013-03475
Event Type
Malfunction
Date Received
June 18, 2013
Date of Event
May 9, 2013
Report Date
May 20, 2013
Manufacturer
SYNTHES GMBH
Product Code
LXT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. THE PFNA-II BLADE SHOWS GRINDING MARKS ON THE OUTSIDE SURFACE AND ON THE TIP ONE EDGE IS BROKEN AWAY. THIS DAMAGE DOES NOT AFFECT THE FUNCTION OF THE BLADE. IT IS POSSIBLE THAT ONE OF THE FOLLOWING POINTS COULD HAVE LEAD TO THE DAMAGES ON THE BLADE: THE BLADE WAS NOT CORRECT MOUNTED ON THE INSERTER. AFTER INSERTION IN TO THE BONE THE INSERTER WAS UNLOCKED OFF THE BLADE TOO EARLY. THE BLADE WAS NOT OPERATED AS DESCRIBED IN THE OPERATION GUIDE. NO PRODUCT FAULT COULD BE DETECTED. PLACEHOLDER.

Additional Manufacturer Narrative · 1

THIS DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES, BUT IS SIMILAR TO DEVICE MARKETED IN THE USA. THE MANUFACTURING RECORDS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND.

Description of Event or Problem · 1

A DEVICE REPORT FROM SYNTHES (B)(4) INDICATED A HOSPITAL IN (B)(6) REPORTED: THE PATIENT HAD A FEMORAL TROCHANTERIC FRACTURE. THE INSERTER DID NOT LOCK THE BLADE. IN THE HOSPITAL WITH OVER 130 OPERATIONS OF JPA, THE ADVERSE EVENT OF UNLOCKING OF THE BLADE HAPPENED. THE SALES REP DID NOT ATTEND THE OPERATION. IT WAS NOTED: THE OPERATION SUCCEEDED BY HAMMERING OF THE BLADE FITTING INTO THE INSERTER, HOWEVER, THE SURGEON FOUND THERE WAS A SPACE IN X-RAY IMAGE WHEN THE BLADE WAS LOCKED, THE INSERTER ROTATED FREE. THE SURGEON COULD CHOOSE THE USAGE OF JPA HOLLOW DRIVER IN ORDER TO LOCK IT; HOWEVER HE USED ANOTHER PRODUCT WITHOUT SUCH A CHOICE. THIS IS 1 OF 1 REPORT FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
275345 PFNA-II BLADE L85 TAN LXT SYNTHES GMBH 8355251

Patients

Seq Age Sex Outcome Treatment
1