FDA Adverse Event Malfunction Summary report: N

ENDOWRIST STABILIZER INSTRUMENT

MDR report key: 3173969 · Received June 17, 2013

Report

Report Number
2955842-2013-02200
Event Type
Malfunction
Date Received
June 17, 2013
Date of Event
December 6, 2012
Report Date
May 21, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K080291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING CONFIRMED THE REPORTED COMPLAINT. THE PITCH CABLE WAS FRAYED. THE PITCH DOWN CABLE WAS FRAYED AT THE DISTAL CLEVIS HUB. THE FRAYED STRANDS STUCK OUT AT THE INSTRUMENT'S WRIST. OTHER CABLES AT THE WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER A DA VINCI S CARDIAC PROCEDURE NURSE CHECKED THE ENDOWRIST STABILIZER INSTRUMENT'S FUNCTION AND FOUND A PART OF THIS PRODUCT WAS BROKEN. THERE WERE NO MISSING PIECES OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
273090 ENDOWRIST STABILIZER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420182-07 S10110613 418

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI S SYSTEM INSTRUMENTS AND ACCESSORIES