FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3173960 · Received June 17, 2013

Report

Report Number
3004209178-2013-94783
Event Type
Malfunction
Date Received
June 17, 2013
Date of Event
June 13, 2013
Report Date
June 13, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED ALL FUNCTIONAL TESTS, INCLUDING PRIME, DISPLACEMENT, REWIND, BASIC OCCLUSION TESTS. UNIT WAS RECEIVED STUCK IN MOTOR ERROR ALARM LOOP DURING BOLUS /BASAL DELIVERY. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND IT PASSED. THE MOTOR MAY HAD AN INTERMITTENT FAILURE THAT IT WAS NOT DETECTED DURING OUR TESTING. UNIT HAD CRACKED RESERVOIR TUBE LIP, BATTERY TUBE THREADS AND CASE NEAR DISPLAY WINDOW CORNERS NOTED DURING VISUAL INSPECTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP ALARMED MOTOR ERROR. IT WAS STATED THAT THE BATTERY WAS REMOVED AND NEW BATTERY WAS INSERTED TO PERFORM THE DISPLACEMENT TEST, BUT THE TEST FAILED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
273087 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-554WWL

Patients

Seq Age Sex Outcome Treatment
1