FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3173956
·
Received June 17, 2013
Report
- Report Number
- 3004209178-2013-94808
- Event Type
- Malfunction
- Date Received
- June 17, 2013
- Date of Event
- April 30, 2013
- Report Date
- May 31, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTONS DUE TO CORRODED KEYPAD TRACES. UNABLE TO CONFIRM THE BUTTON ERROR ALARM. THE DEVICE WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A LOOSE/FLUSH DRIVE SUPPORT DISK. THE UNIT WAS RECEIVED WITH CRACKED CASE AT THE DISPLAY WINDOW CORNERS, BATTERY TUBE THREADS, AND SCRATCHED SCREEN.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP HAD UNRESPONSIVE BUTTONS AND ALARMED BUTTON ERROR. THE CUSTOMER'S BLOOD GLUCOSE READING WAS 10.0MMOL/L. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.