ENDURANT II
Report
- Report Number
- 2953200-2013-01138
- Event Type
- Injury
- Date Received
- June 17, 2013
- Date of Event
- May 22, 2013
- Report Date
- May 23, 2013
- Manufacturer
- MEDTRONIC IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P100021
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). METHODS: FILMS. RESULTS: NARROWED DISTAL LEFT ILIAC. STENT GRAFT OCCLUSION. LANDING IN THE NARROWED DISTAL LEFT ILIAC. CONCLUSIONS: NARROWED DISTAL LEFT ILIAC. STENT GRAFT OCCLUSION. LANDING IN THE NARROWED DISTAL LEFT ILIAC.
AN ENDURANT II STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 50.5 MM IN DIAMETER ABDOMINAL AORTIC ANEURYSM. THE PROXIMAL AORTIC NECK WAS 25.7-28.5 MM IN DIAMETER AND 33 MM IN LENGTH. THE DISTAL AORTA WAS 46MM IN DIAMETER. THE LEFT ILIAC ARTERY DIAMETER WAS 18MM PROXIMALLY AND 15MM DISTALLY. THE LEFT FEMORAL ARTERY WAS 8MM IN DIAMETER. THE RIGHT ILIAC ARTERY DIAMETERS WERE 36-29-23.5MM. THE RIGHT FEMORAL ARTERY DIAMETER WAS 9MM. THE RIGHT ILIAC ARTERY WAS SEVERELY CALCIFIED, AND THE LEFT ILIAC ARTERY WAS MILDLY CALCIFIED. THE RIGHT INTERNAL ILIAC ARTERY WAS INTENTIONALLY COVERED AND COILED DURING THE IMPLANT. IT WAS REPORTED THAT APPROXIMATELY ONE WEEK POST IMPLANT THE LEFT LIMB WAS FOUND TO BE OCCLUDED WITH THROMBUS. A THROMBECTOMY AND STENTING WERE PERFORMED, AND THE OCCLUSION RESOLVED. THE PHYSICIAN NOTED THAT ALTHOUGH THE ORIGINAL FILMS LOOKED OKAY, THERE WAS COMPRESSION OF THE DISTAL PORTION OF THE LEFT LIMB. THIS WAS DUE TO THE 20MM FLARED END LANDING IN THE NARROWED DISTAL LEFT ILIAC. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PATIENT IS FINE. A REVIEW OF RETURNED FILMS PRE-IMPLANT SHOWED A 5CM AAA WHICH WAS CALCIFIED AND CONTAINED THROMBUS. THE RIGHT COMMON ILIAC WAS ANEURYSMAL; MEASURED 3.4CM MAX DIAMETER. THE LEFT COMMON ILIAC DIAMETER MEASURED APPROXIMATELY 18-19MM AT THE ORIGIN, NARROWING TO 14-15MM AT THE DISTAL END, AND WAS CALCIFIED. FILM POST-STENT GRAFT IMPLANT WERE NOT PROVIDED; THEREFORE THE REPORTED OCCLUSION COULD NOT BE ASSESSED. IT IS POSSIBLE THAT PLACEMENT OF THE 20MM STENT GRAFT INTO THE DISTAL NARROWED PORTION OF THE CALCIFIED LEFT COMMON ILIAC MAY HAVE CONTRIBUTED TO THE OCCLUSION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 274406 | ENDURANT II | SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | MEDTRONIC IRELAND | V04005949 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00068 YR | Required Intervention |