ASR UNI FEMORAL IMPL SIZE 46
Report
- Report Number
- 1818910-2013-19065
- Event Type
- Injury
- Date Received
- June 17, 2013
- Date of Event
- May 13, 2014
- Report Date
- October 24, 2014
- Manufacturer
- DEPUY INTL., LTD. - 8010379
- Product Code
- KWA
- PMA / PMN Number
- PK040627
- Removal / Correction Number
- Z-1749/1816-2011
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- ATTORNEY
Narratives
THERE IS NO NEW INFORMATION THAT WOULD CHANGE THE OUTCOME OF THE INVESTIGATION. THE COMPLAINT IS CONSIDERED CLOSED.
THE ASR PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET IN AUGUST 2010, AND THE ASR PRODUCT CODES ARE NOW CONSIDERED INACTIVE. FURTHER INVESTIGATION OF THIS INDIVIDUAL INCIDENT WILL NOT BE UNDERTAKEN, AS THERE IS AN ONGOING INVESTIGATION REGARDING THE ROOT CAUSE(S) AND/OR CORRECTIVE ACTIONS. REF. (B)(4). DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.
IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. DEPUY STILL CONSIDERS THIS INVESTIGATION CLOSED.
IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. DEPUY STILL CONSIDERS THE INVESTIGATION CLOSED.
UPDATE REC'D 05/13/2014 - SALES REP REPORTED REVISION SURGERY. PATIENT WAS REVISED TO ADDRESS OSTEOLYSIS. THERE IS NO NEW ADDITIONAL INFORMATION THAT WOULD AFFECT THE INVESTIGATION. THIS COMPLAINT WAS UPDATED ON: 06/06/2014.
UPDATE REC'D 10/24/2014 MEDICAL RECORDS RECEIVED. PATIENT WAS REVISED TO ADDRESS METALLOSIS. UPON REVISION, GRAY STAINED FLUID COLLECTION, GRAY STAINED TISSUE, AND CORROSION ON THE TRUNNION WERE NOTED. THE STEM AND ASR SLEEVE ARE BEING ADDED TO THE COMPLAINT.
LITIGATION ALLEGES INTERMITTENT CLICKING, PAIN, STIFFNESS, DISCOMFORT AND ELEVATED METAL ION LEVELS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 274364 | ASR UNI FEMORAL IMPL SIZE 46 | TOTAL HIP REPLACEMENT | KWA | DEPUY INTL., LTD. - 8010379 | 2716546 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Other |