COULTER® AC T 5DIFF AL ANALYZER
Report
- Report Number
- 1061932-2013-01112
- Event Type
- Malfunction
- Date Received
- June 17, 2013
- Date of Event
- May 22, 2013
- Report Date
- May 22, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K030291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE FSE INSPECTED THE INSTRUMENT AND CONFIRMED THE HEMOGLOBIN (HGB) LYSE LEAKED FROM THE HGB REAGENT SYRINGE. THE FSE REPLACED THE HGB REAGENT SYRINGE ASSEMBLY, RESOLVING THE REPORTED ISSUE. THE FSE PRIMED ALL REAGENTS, RAN A STARTUP, AND LATEX CHECKS, RAN ALL QUALITY CONTROL (QC) AND PATIENT SAMPLES. THE FSE VERIFIED THE SYSTEM WAS OPERATIONAL. THE FAILURE MODE IDENTIFIED WAS RELATED TO THE HGB REAGENT SYRINGE ASSEMBLY WHICH NEEDED REPLACEMENT. (B)(4).
THE CUSTOMER REPORTED TO BECKMAN COULTER (BEC) A LEAK OF 200 ML OF HEMOGLOBIN (HGB) LYSE REAGENT WHILE THE COULTER AC T 5DIFF AL ANALYZER WAS IN SHUTDOWN. THE LEAK WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE OPERATOR WAS WEARING GLOVES AND A LABORATORY COAT AT THE TIME OF THE INCIDENT AND VERIFIED THAT NO LINES WERE DISCONNECTED. THERE WAS NO BIOHAZARD EXPOSURE TO OPEN WOUNDS OR MUCOUS MEMBRANES. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION WITH THIS EVENT. NO DEATH OR INJURY IS ATTRIBUTED TO THIS EVENT AND THERE WAS NO CHANGE TO PATIENT TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 273516 | COULTER® AC T 5DIFF AL ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | ACT 5DIFF AL | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |