FDA Adverse Event Malfunction Summary report: N

ACT 5DIFF RINSE

MDR report key: 3173829 · Received June 17, 2013

Report

Report Number
1061932-2013-01177
Event Type
Malfunction
Date Received
June 17, 2013
Date of Event
May 15, 2013
Report Date
May 20, 2013
Manufacturer
BECKMAN COULTER
Product Code
JCB
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE FSE INSPECTED AND REPLACED APERTURE, REPLACED O-RINGS, AND WHITE BLOOD CELL (WBC) BATH. UPON FURTHER INSPECTION, THE FSE OBSERVED MUCUS LIKE SUBSTANCE ON AN OLD ELECTRODE THAT HAD JUST BEEN REMOVED. THE FSE RAN SEVERAL SAMPLES AND THE MUCUS SUBSTANCE APPEARED ON THE WBC ELECTRODE AGAIN. THE FSE DISCOVERED THE MUCUS SUBSTANCE WAS IN THE RINSE REAGENT BOTTLE. PER PHONE CONVERSATION WITH THE CUSTOMER, THE SUBJECT REAGENT HAD BEEN IN USE FOR 11 DAYS WHEN THE ISSUE WAS OBSERVED. NO RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY; SAMPLES WERE SENT OUT FOR ANALYSIS. THE FSE DECONTAMINATED THE INSTRUMENT AND REPLACED THE RINSE REAGENT WITH A NEW BOTTLE, EXPIRATION DATE 03/20/2014, AND VERIFIED THE INSTRUMENT'S OPERATION. FAILURE MODE WAS ATTRIBUTED TO THE MUCUS LIKE SUBSTANCE PRESENT IN THE RINSE REAGENT BOTTLE. IT IS UNKNOWN HOW THE REAGENT GOT CONTAMINATED. THE ANALYZER GENERATED MESSAGES FOR SOME PARAMETERS WHICH ALERTED THE CUSTOMER TO REVIEW THE RESULTS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO BECKMAN COULTER (BEC) THAT THE COULTER AC T 5DIFF CP ANALYZER WAS GIVING WHITE BLOOD COUNT (WBC) *, VOTING OUT, AND GIVING NO NUMERICAL WBC RESULTS ON PATIENT SAMPLES. THE CUSTOMER INDICATED THAT PATIENT #5 WAS TESTED TWICE ON THE AC T 5DIFF CP ANALYZER AND ERRONEOUS RESULTS WERE OBTAINED. CORRECT RESULTS WERE NOT PROVIDED BY THE CUSTOMER FOR THIS PATIENT AND IT IS UNKNOWN IF THE PATIENT SAMPLE WAS TESTED AT THE HOSPITAL'S LABORATORY. PATIENT'S RESULTS PROVIDED BY THE CUSTOMER ARE SUPPLIED IN THE FILE ATTACHMENT. THE ERRONEOUS RESULTS WERE NOT REPORTED OUTSIDE OF THE LABORATORY. THERE WAS NO DEATH, INJURY OR EFFECT TO PATIENT TREATMENT. THREE (3) MDRS ARE BEING SUBMITTED FOR THE ERRONEOUS CBC/DIFF RESULTS OBTAINED AT THIS CUSTOMER SITE: 1061932-2013-01169, 1061932-2013-01170, 1061932-2013-01177.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
273341 ACT 5DIFF RINSE DETERGENT JCB BECKMAN COULTER NA 13002C00424

Patients

Seq Age Sex Outcome Treatment
1 80 YR