Report Number
0001825034-2013-01994
Event Type
Injury
Date Received
June 17, 2013
Date of Event
September 6, 2012
Report Date
October 18, 2013
Manufacturer
BIOMET ORTHOPEDICS
Product Code
KWA
PMA / PMN Number
PK042037
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING FILED TO RELAY THE CORRECT NOTIFICATION DATE.

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING FILED TO RELAY ADDITIONAL INFORMATION, WHICH WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH.

Additional Manufacturer Narrative · 1

CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 1 STATES, "MATERIAL SENSITIVITY REACTIONS." NUMBER 6 STATES, "INADEQUATE RANGE OF MOTION DUE TO IMPROPER SELECTION OR POSITIONING OF COMPONENTS." NUMBER 14 STATES, "INTRAOPERATIVE OR POSTOPERATIVE BONE FRACTURE AND/OR POSTOPERATIVE PAIN." NUMBER15 STATES, "ELEVATED METAL ION LEVELS HAVE BEEN REPORTED WITH METAL-ON-METAL ARTICULATING SURFACES." THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT, AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. THIS REPORT IS NUMBER 2 OF 3 MDRS FILED FOR THE SAME EVENT (REFERENCE 1825034-2013-01993 / 01995).

Description of Event or Problem · 1

PATIENT'S LEGAL COUNSEL REPORTED THAT PATIENT UNDERWENT TOTAL HIP ARTHROPLASTY ON (B)(6)2010. SUBSEQUENTLY, LEGAL COUNSEL FOR PATIENT REPORTS PATIENT WAS REVISED ON (B)(6) 2012, DUE TO PATIENT ALLEGATIONS OF PAIN, DISCOMFORT, SORENESS, DYSFUNCTION, LOSS OF RANGE OF MOTION, ELEVATED METAL ION LEVELS, SWELLING, INFLAMMATION, DAMAGE TO SURROUNDING BONE AND TISSUE, METALLOSIS, METAL POISONING, AND LACK OF MOBILITY. NO FURTHER INFORMATION HAS BEEN PROVIDED TO DATE. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.

Description of Event or Problem · 1

PATIENT'S LEGAL COUNSEL REPORTED THAT PATIENT UNDERWENT TOTAL HIP ARTHROPLASTY ON (B)(6) 2010. SUBSEQUENTLY, LEGAL COUNSEL FOR PATIENT REPORTS PATIENT WAS REVISED ON (B)(6) 2012 DUE TO PATIENT ALLEGATIONS OF PAIN, DISCOMFORT, SORENESS, DYSFUNCTION, LOSS OF RANGE OF MOTION, ELEVATED METAL ION LEVELS, SWELLING, INFLAMMATION, DAMAGE TO SURROUNDING BONE AND TISSUE, METALLOSIS, METAL POISONING, AND LACK OF MOBILITY. ADDITIONAL INFORMATION RECEIVED IN PATIENT'S MEDICAL RECORDS INDICATES THAT REVISION PROCEDURE ON (B)(6) 2012 WAS DUE TO PAIN, BROWN FLUID, PSEUDOTUMOR, AND INFLAMED BROWN SOFT TISSUE. THE ACETABULAR CUP, TAPER ADAPTER AND HEAD WERE REMOVED AND REPLACED. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
273657 M2A-MAGNUM 42-50MM TPR INSRT-6 PROSTHESIS, HIP KWA BIOMET ORTHOPEDICS N/A 377090

Patients

Seq Age Sex Outcome Treatment
1 43 YR Hospitalization| R