ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-08360
- Event Type
- Malfunction
- Date Received
- June 17, 2013
- Date of Event
- May 21, 2013
- Report Date
- May 21, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/12/2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE PEELING AROUND THE UP ARROW KEYPAD BUTTON. A DAMAGED KEYPAD WILL PERMIT CONTAMINATION TO PERMEATE THE BUTTONS WHICH WILL HAVE A NEGATIVE IMPACT ON BUTTON FUNCTION. THIS SITUATION IS NOT LIKELY TO RESULT IN AN ADVERSE EVENT AS THE DAMAGED KEYPAD SHOULD BE CLEARLY VISIBLE AND WARNS THE PATIENT TO DISCONTINUE USING THE PUMP. THE OWNER'S BOOKLET INSTRUCTS THE USER TO CALL ANIMAS CUSTOMER SERVICE IF THE USER SUSPECTS THAT THE PRODUCT IS DAMAGED. THE OK, UP ARROW AND CONTRAST KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONSIVE. THE DOWN ARROW BUTTON WAS FOUND TO BE RESPONSIVE. THE KEYPAD COVER REMOVED AND FOUND CONTAMINATION UNDER ALL BUTTON CONTACTS. UNRELATED THE COMPLAINT, THE AUDIBLE SOUND WAS UNABLE TO BE TESTED DUE TO A DETACHED BOLUS BUTTON KEYPAD COVER.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE WITH THE UP/DOWN/OK BUTTONS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.