HOMECHOICE
Report
- Report Number
- 1416980-2013-15610
- Event Type
- Malfunction
- Date Received
- June 17, 2013
- Date of Event
- April 23, 2013
- Report Date
- May 24, 2013
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION: THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MIN DRAIN VOLUME THRESHOLD. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.
DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 00:25:34. DURING NIGHT DRAIN CYCLE SEVEN, THE PATIENT'S ULTRAFILTRATION READING WAS 914ML, INDICATING THE HOME PATIENT (HP) DRAINED 914ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1500ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 274652 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |