FDA Adverse Event Malfunction Summary report: N

ENTERALITE INFINITY ENTERAL FEEDING PUMP

MDR report key: 3173493 · Received June 12, 2013

Report

Report Number
1722139-2013-01895
Event Type
Malfunction
Date Received
June 12, 2013
Date of Event
February 15, 2013
Report Date
May 17, 2013
Manufacturer
MOOG MEDICAL DEVICES GROUP
Product Code
LZH
PMA / PMN Number
K031199
Removal / Correction Number
Z-0294-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION FOUND THAT THE DIODE D7 WAS DETERMINED TO BE LEAKY AND WAS OUT OF THE SPECIFICATION. NEW PCB BOARD WAS REPLACED TO SOLVE THE ISSUE. REFERENCE RECALL NUMBER: Z-0294-2013.

Description of Event or Problem · 1

CUSTOMER STATED THAT PUMP ALARMED ERROR CODE 13 WHEN TURNED ON. NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
266321 ENTERALITE INFINITY ENTERAL FEEDING PUMP LZH MOOG MEDICAL DEVICES GROUP INFINITY PUMP

Patients

Seq Age Sex Outcome Treatment
1