FUSION OMNI-TOME
Report
- Report Number
- 1037905-2013-00350
- Event Type
- Malfunction
- Date Received
- June 6, 2013
- Date of Event
- May 8, 2013
- Report Date
- May 9, 2013
- Manufacturer
- WILSON-COOK MEDICAL INC.
- Product Code
- KNS
- PMA / PMN Number
- K052051
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
INVESTIGATION EVALUATION: OUR LABORATORY EVALUATION OF THE PRODUCT SAID TO BE INVOLVED CONFIRMED THE REPORT. HOWEVER, DUE TO THE CONDITION OF THE DEVICE A FUNCTIONAL TEST FOR ORIENTATION WAS NOT CONDUCTED. THE CUTTING WIRE WAS OBSERVED EXTENDING FROM THE DISTAL END OF THE CATHETER APPROXIMATELY 26CM. THE CUTTING WIRE HAS DISCONNECTED FROM THE DRIVE WIRE AT THE CENTER CANNULA JOINT, APPROXIMATELY 36.5CM FROM THE DISTAL END OF THE SPHINCTEROTOME. THE CENTER CANNULA JOINT EXHIBITS A BREAK. THE AREA OF THE BREAK AND DISCONNECTION IS LOCATED INSIDE THE OUTER CATHETER. ALSO, THE CUTTING WIRE SECURING COMPONENT LOCATED NEAR THE DISTAL END OF THE SPHINCTEROTOME HAS DISCONNECTED FROM THE CATHETER TIP. THIS DISCONNECTION ALLOWED A PORTION OF THE CUTTING WIRE TO MOVE FORWARD EXTENDING OUTSIDE THE CATHETER TIP. NO PORTION OF THE SECURING COMPONENT IS MISSING. THE SKIVE AREA WHERE THE CUTTING WIRE SECURING COMPONENT IS POSITIONED INSIDE THE CATHETER TIP IS TORN. THE CUTTING WIRE DOES NOT EXHIBIT EVIDENCE OF A CAUTERY APPLICATION (BLACKENING OF THE CUTTING WIRE WAS NOT OBSERVED). THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED BECAUSE THE CONDITION OF THE PRODUCT SAID TO BE INVOLVED PROHIBITED A COMPLETED EVALUATION. A DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THE REPORTED OBSERVATION WAS NOT OBSERVED DURING OUR LABORATORY ANALYSIS OF THE RETURNED PRODUCT. IMPROPER CUTTING WIRE ORIENTATION CAN OCCUR IF THE DISTAL END OF THE CATHETER IS SHAPED MANUALLY. THIS SPHINCTEROTOME CATHETER IS PRECURVED AND IS PROVIDED WITH A PRECURVED STYLET IN THE DISTAL TIP OF THE CATHETER. THIS OBVIATES THE NEED FOR MANUAL FORMATION. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT APPLY MANUAL PRESSURE TO TIP OR CUTTING WIRE OF THE SPHINCTEROTOME TO INFLUENCE ORIENTATION, AS THIS MAY RESULT IN DAMAGE TO DEVICE." OTHER FACTORS THAT CAN CONTRIBUTE TO IMPROPER CUTTING WIRE ORIENTATION INCLUDE MANIPULATING THE HANDLE WITH THE CATHETER IN A COILED POSITION OR WITH THE PRECURVED STYLET INSIDE THE CANNULATING TIP. THE INSTRUCTIONS FOR USE ADVISE THE USER TO UNCOIL AND STRAIGHTEN THE SPHINCTEROTOME UPON REMOVING THE DEVICE FROM THE PACKAGING. THE USER IS THEN INSTRUCTED TO CAREFULLY REMOVE THE PRECURVED STYLET FROM THE CANNULATING TIP. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT EXERCISE HANDLE WHILE DEVICE IS COILED OR PRECURVED STYLET IS IN PLACE, AS THIS MAY CAUSE DAMAGE TO SPHINCTEROTOME AND RENDER IT INOPERABLE." BROKEN AT CENTER CANNULA: SEPARATION OF THE DRIVE WIRE FROM THE CUTTING WIRE CAN OCCUR IF THE TIP OF THE DEVICE IS OVER FLEXED. THE INSTRUCTIONS FOR USE CAUTION THE USER NOT TO OVER FLEX OR BOW THE TIP BEYOND 90 DEGREES, AS THIS MAY DAMAGE THE SPHINCTEROTOME. THE INSTRUCTIONS FOR USE CAUTION THE USER THE ELEVATOR SHOULD REMAIN OPEN/DOWN WHEN ADVANCING OR RETRACTING THE SPHINCTEROTOME. IF THE ELEVATOR OF THE ENDOSCOPE IS PLACED IN THE CLOSED/UP POSITION WITH THE SPHINCTEROTOME INSIDE THE ACCESSORY CHANNEL, THIS COULD CONSTRICT MOVEMENT OF THE DRIVE WIRE WHEN THE HANDLE IS MANIPULATED IN AN ATTEMPT TO BOW THE SPHINCTEROTOME. IF ADDED PRESSURE IS APPLIED TO THE HANDLE WITH THE ELEVATOR AND SPHINCTEROTOME IN THIS POSITION, THIS COULD CONTRIBUTE TO DRIVE WIRE SEPARATION FROM THE CUTTING WIRE. OTHER FACTORS THAT CAN CONTRIBUTE TO DRIVE WIRE SEPARATION INCLUDE MANIPULATING THE HANDLE WITH THE CATHETER IN A COILED POSITION OR WITH THE PRECURVED STYLET INSIDE THE CANNULATING TIP. THE INSTRUCTIONS FOR USE ADVISE THE USER TO UNCOIL AND STRAIGHTEN THE SPHINCTEROTOME UPON REMOVING THE DEVICE FROM THE PACKAGING. THE USER IS THEN INSTRUCTED TO CAREFULLY REMOVE THE PRECURVED STYLET FROM THE CANNULATING TIP. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT EXERCISE HANDLE WHILE DEVICE IS COILED OR PRECURVED STYLET IS IN PLACE, AS THIS MAY CAUSE DAMAGE TO SPHINCTEROTOME AND RENDER IT INOPERABLE." CUTTING WIRE SECURING COMPONENT SEPARATION: SEPARATION OF THE CUTTING WIRE SECURING COMPONENT AND THE CATHETER CAN OCCUR IF THE TIP OF THE SPHINCTEROTOME IS OVER FLEXED. THE INSTRUCTIONS FOR USE CAUTION THE USER NOT TO OVER FLEX OR BOW THE TIP BEYOND 90 DEGREES, AS THIS MAY DAMAGE THE SPHINCTEROTOME. THIS TYPE OF DAMAGE CAN OCCUR IF THE DISTAL END OF THE CATHETER IS SHAPED MANUALLY. THIS SPHINCTEROTOME CATHETER IS PRE-CURVED AND IS PROVIDED WITH A PRE-CURVED STYLET IN THE DISTAL TIP OF THE CATHETER. THIS OBVIATES THE NEED FOR MANUAL FORMATION. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT APPLY MANUAL PRESSURE TO TIP OR CUTTING WIRE OF THE SPHINCTEROTOME TO INFLUENCE ORIENTATION, AS THIS MAY RESULT IN DAMAGE TO DEVICE." IF THE ELEVATOR OF THE ENDOSCOPE REMAINS IN THE CLOSED/UP POSITION WHEN RETRACTION OF THE SPHINCTEROTOME IS ATTEMPTED AND ADDITIONAL PRESSURE IS APPLIED, THIS COULD HAVE CONTRIBUTED TO SEPARATION OF THE CATHETER AND CUTTING WIRE SECURING COMPONENT. THE INSTRUCTIONS FOR USE CAUTION THE USER THE ELEVATOR SHOULD REMAIN OPEN/DOWN WHEN ADVANCING OR RETRACTING THE SPHINCTEROTOME. THIS ACTIVITY WILL AID IN DEVICE PRESERVATION. OTHER FACTORS THAT CAN CONTRIBUTE TO SEPARATION OF THE CUTTING WIRE SECURING COMPONENT AND THE CATHETER INCLUDE MANIPULATING THE HANDLE WITH THE CATHETER IN A COILED POSITION OR WITH THE PRECURVED STYLET INSIDE THE CANNULATING TIP. THE INSTRUCTIONS FOR USE ADVISE THE USER TO UNCOIL AND STRAIGHTEN THE SPHINCTEROTOME UPON REMOVING THE DEVICE FROM THE PACKAGING. THE USER IS THEN INSTRUCTED TO CAREFULLY REMOVE THE PRECURVED STYLET FROM THE CANNULATING TIP. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT EXCISE HANDLE WHILE DEVICE IS COILED OR PRECURVED STYLET IS IN PLACE, AS THIS MAY CAUSE DAMAGE TO SPHINCTEROTOME AND RENDER IT INOPERABLE." PRIOR TO DISTRIBUTION, ALL FUSION OMNI-TOME SPHINCTEROTOMES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TEST TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SID TO BE INVOLVED MET ALL MFG REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK FUSION OMNI-TOME. THE CUTTING WIRE ORIENTATION OF THE SPHINCTEROTOME WAS REPORTED TO BE OFF. SEE MDR 1037905-2013-00349. THE PHYSICIAN REMOVED THE DEVICE FROM THE ENDOSCOPE AND PROCEEDED BY USING A SECOND COOK FUSION OMNI-TOME. THE ORIENTATION OF THE SECOND DEVICE WAS REPORTED TO BE OFF AS WELL. THE PHYSICIAN HAD THE NURSE ROTATE THE HANDLE TO CORRECT THE ORIENTATION. THEY WERE ABLE TO PERFORM A SPHINCTEROTOMY AND COMPLETE THE PROCEDURE USING THE SECOND DEVICE. THEY REMOVED THE DEVICE FROM THE ENDOSCOPE AND ALSO FOUND THE CUTTING WIRE HAD BROKEN BUT WAS STILL ATTACHED. NO SECTION OF THE DEVICE DETACHED INSIDE THE PT. SEE MDR 1037905-2013-00350. A SECOND OF THE DEVICE DID NOT REMAIN INSIDE THE PT'S BODY. THE PT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 251855 | FUSION OMNI-TOME | KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) | KNS | WILSON-COOK MEDICAL INC. | W3269482 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | COOK METRO II WIRE GUIDE| OLYMPUS V-ENDOSCOPE (UNK MODEL NUMBER) |